Secure Workstation for OpenLab CDS
ChemStation Edition
Installation Guide
Secure Workstation for OpenLab CDS ChemStation Edition Installation
Secure Workstation for OpenLab CDS ChemStation Edition Installation
Notices
Document Information
Part No: M8313-90004 Rev. D
EDITION 05/2020
Copyright
© Agilent Technologies, Inc. 2014-2020
No part of this manual may be repro-
duced in any form or by any means
(including electronic storage and retrieval
or translation into a foreign language)
without prior agreement and written con-
sent from Agilent Technologies, Inc. as
governed by United States and interna-
tional copyright laws.
Agilent Technologies
Hewlett-Packard-Strasse 8
76337 Waldbronn
Warranty
The material contained in this document
is providedas is,and is subject to being
changed, without notice, in future edi-
tions. Further, to the maximum extent
permitted by applicable law, Agilent dis-
claims all warranties, either express or
implied, with regard to this manual and
any information contained herein, includ-
ing but not limited to the implied warran-
ties of merchantability and fitness for a
particular purpose. Agilent shall not be lia-
ble for errors or for incidental or conse-
quential damages in connection with the
furnishing, use, or performance of this
document or of any information con-
tained herein. Should Agilent and the user
have a separate written agreement with
warranty terms covering the material in
this document that conflict with these
terms, the warranty terms in the separate
agreement shall control.
Technology Licenses
The hardware and/or software described
in this document are furnished under a
license and may be used or copied only in
accordance with the terms of such
license.
Restricted Rights Legend
U.S. Government Restricted Rights. Soft-
ware and technical data rights granted to
the federal government include only
those rights customarily provided to end
user customers. Agilent provides this
customary commercial license in Soft-
ware and technical data pursuant to FAR
12.211 (Technical Data) and 12.212
(Computer Software) and, for the Depart-
ment of Defense, DFARS 252.227-7015
(Technical Data - Commercial Items) and
DFARS 227.7202-3 (Rights in Commer-
cial Computer Software or Computer
Software Documentation).
Safety Notices
CAUTI ON
A CAUTION notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly per-
formed or adhered to, could result
in damage to the product or loss
of important data. Do not proceed
beyond a
CAUTION notice until
the indicated conditions are fully
understood and met.
WARNING
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly per-
formed or adhered to, could
result in personal injury or death.
Do not proceed beyond a WARN-
ING notice until the indicated
conditions are fully understood
and met.
Software Revision
This guide is valid for Secure
Workstation for OpenLab CDS
ChemStation Edition revision C.01.10.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 3
In this Guide ...
This installation guide provides instructions to install and configure Agilent Secure
Workstation for OpenLab CDS ChemStation Edition.
1 Prepare your PC
This chapter describes how to configure a non-Agilent-delivered PC.
Agilent-delivered PC Bundle systems are delivered with the supported pre-installed
Windows operating system and are configured for optimum performance.
Non-Agilent PCs require some manual configuration changes in order to provide
optimum performance.
2 Install the Software
The installation is automated by the OpenLab CDS ChemStation Edition Installer.
This tool installs all components needed.
3 Post Installation Tasks
This chapter describes tasks that are relevant after finishing the installation.
4 Optional Procedures
This chapter contains information on the Additional Drivers and Software wizard
and other helpful procedures.
Table 1 Terms and abbreviations used in this document
Term Description
AIC Agilent's Analytical Instrument Controller
CDS Chromatography Data System
ChemStation OpenLab CDS ChemStation Edition
Control Panel OpenLab Control Panel
Microsoft Control Panel Part of the Microsoft Windows operating system
Secure Workstation Secure Workstation for OpenLab CDS ChemSta-
tion Edition
Secure Workstation for OpenLab CDS ChemStation Edition Installation 4
5 Secure File I/O
This chapter describes how to protect your ChemStation folders with Secure File
I/O.
6 Secure Workstation for OpenLab CDS ChemStation Edition Licensing
This chapter describes the licensing of a Secure Workstation for OpenLab CDS
ChemStation Edition.
7 Configure Secure Workstation for OpenLab CDS ChemStation Edition
This chapter describes the configuration of Secure Workstation for OpenLab CDS
ChemStation Edition.
8 Upgrade Secure Workstation
This chapter describes the upgrade of Secure Workstation for OpenLab CDS
ChemStation Edition.
9 Uninstall the Software
This chapter contains information on the uninstallation by using the OpenLab
Uninstallation Wizard.
10 Troubleshooting
The chapter gives some troubleshooting hints.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 5
Contents
1 Prepare your PC 7
Install and Update Windows 8
Run the Site Preparation Tool 10
Install and Configure Third Party Tools 11
Configure a Printer 16
Before You Begin the Installation of ChemStation 18
Configure the Operating System 20
2 Install the Software 31
Prepare for Installation 32
Run the OpenLab CDS Installation Wizard 33
3 Post Installation Tasks 36
Run an Software Verification after Software Installation 37
Allow ChemMain Through Firewall 38
Configure the Antivirus Program 39
4 Optional Procedures 41
Install Additional Software and Drivers 42
Copy installation Files to a Centralized Folder for Installation 43
Add Shortcut to Public Folder 44
5 Secure File I/O 45
About Secure File I/O 46
Prepare Secure File I/O in Windows 47
Enable Secure File I/O in ChemStation 48
Check privileges for ChemStation users 49
Important Notes for Working with Secure File I/O 50
Limitations 51
Secure Workstation for OpenLab CDS ChemStation Edition Installation 6
6 Secure Workstation for OpenLab CDS ChemStation Edition
Licensing 52
About OpenLab Licensing 53
Obtain your Software License 54
7 Configure Secure Workstation for OpenLab CDS ChemStation
Edition 59
First Login 60
Add your License to the Control Panel 61
Add Users 62
Add Users to a Role 63
8 Upgrade Secure Workstation 64
Planning your Upgrade 65
Convert Classic Instrument Drivers to RC.NET 66
Preparations for the Upgrade 69
License Upgrade 71
Upgrade a Secure Workstation 73
9 Uninstall the Software 75
About Uninstallation 76
Run the OpenLab CDS Uninstallation Wizard 77
10 Troubleshooting 78
Reconfigure Instruments Using Classic Drivers After Upgrade to
ChemStation C.01.08 or Higher 79
11 Appendix 82
Change the PC Name 83
Sales and Support Assistance 84
Secure Workstation for OpenLab CDS ChemStation Edition Installation 7
1 Prepare your PC
Install and Update Windows 8
Run the Site Preparation Tool 10
Install and Configure Third Party Tools 11
Install and Configure Adobe Acrobat Reader 11
Install Keysight IO Libraries Suite 13
Install the .NET Framework 14
Configure a Printer 16
Add a Network Printer as a Local Printer 17
Before You Begin the Installation of ChemStation 18
Configure the Operating System 20
PC Configuration Checker 20
Manual Configuration Steps 21
Edit the Security Settings for LC/MS and CE/MS Systems 29
This chapter describes how to configure a non-Agilent-delivered PC.
Agilent-delivered PC Bundle systems are delivered with the supported
pre-installed Windows operating system and are configured for optimum
performance. Non-Agilent PCs require some manual configuration changes in
order to provide optimum performance.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 8
1 Prepare your PC
Install and Update Windows
Install and Update Windows
1 Install the Windows operating system from the Microsoft installation media
or qualified PC image media provided by your IT department. During the
setup, provide the computer name, administrator password and network
settings. Choose to either join an existing domain or set up the system in a
workgroup mode.
2 For Windows 10 Pro users: Update to the latest Windows 10 edition in
accordance to the guidelines of your local IT department.
3 To secure your system against viruses please install an antivirus program. Be
sure to open the firewall ports listed in the Firewall Settings in the OpenLab
CDS ChemStation Edition Requirements guide
(CDS_CS_HW-SW-Requirements.pdf), including the ports listed for OpenLab
Server in the tables 14 and 15..
4 In the Microsoft Control Panel under System> Windows activation, click
Change product key. Enter a valid value to activate Windows.
5 Under Windows Update, click Check for updates to check for updates and apply
all critical security patches. Make sure that all Windows updates have been
performed before installing ChemStation (update settings may be configured
centrally by your IT department).
6 Disable or defer Windows updates (update settings may be configured
centrally by your IT department):
Windows 7:
a In the Microsoft Control Panel, select Windows Update.
b Select Change settings.
NOTE
Running antivirus programs might influence the behavior and performance of
your computer. Some virus scanners might cause issues when used with
OpenLab CDS. The application is tested with Symantec Endpoint Protection 14.0
MP2 and with Microsoft Security Essentials.
NOTE
This setting is important to avoid data loss due to system reboot during data
acquisition.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 9
1 Prepare your PC
Install and Update Windows
c In the Important updates section, select Never check for updates. Clear
the other update options.
d Restart the computer after update.
Windows 10:
a Under Settings> Update & Security, select Advanced options.
b Enter the number of days by which the update should be deferred.
7 Windows 10 and Windows 7 only: To configure remote settings: In the
Microsoft Control Panel navigate to System> Remote settings. On the Remote
tab:
In the Remote Assistance section, clear the check box Allow Remote
Assistance connections to this computer.
In the Remote Desktop section, select Don't allow connections to this
computer.
8 In the Microsoft Control Panel under Date and Time: Choose the time zone of
your regional location.
9 In the Microsoft Control Panel under Region (Region and Language for
Windows 7):
a Regional format options should be set to English (United States) from the
drop-down list.
b If regional format other than English (United States) is used, the following
settings are mandatory. The settings can be defined by clicking on the
Additional settings... button:
Decimal symbol = . (point)
Digit grouping symbol = , (comma)
List separator =, (comma)
10 In the Microsoft Control Panel under Region (Region and Language for
Windows 7), on the Administrative tab:
a In the Language for non-Unicode program section, click Change system
locale....
b From the drop down list, select English (United States).
NOTE
Do not change the system locale if you are using an English, Japanese or
Chinese Operating System.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 10
1 Prepare your PC
Run the Site Preparation Tool
Run the Site Preparation Tool
1 Run the installer from the USB medium or from a centralized folder. From the
Planning screen, select System Configuration Checker.
2 The Site Preparation Tool opens. From the drop-down list, select: Secure
Workstation for OpenLab CDS ChemStation Edition.
3 Select OK.
4 Complete page 1 of the Contact InformationSystem details by typing in the
fields provided.
System Location fields
System Information fields
Configuration fields
5 Review the system details and make any necessary entries. The system will
follow the paths specified.
6 Select the green check mark icon in the top left corner of the screen to begin
the software check. A summary report is displayed showing the results for
each check category. Results are expressed as Pass, Warning, Critical Warning,
or Fail.
Fail results must be corrected before continuing with the installation. Agilent
recommends investigating and correcting any Critical Warnings and Warnings
whenever possible before proceeding.
7 To view details of the report, select the appropriate link: System Hardware
Details, Operating System and Software Details, or Manual Verification Required.
8 To save the report, select the Save icon at the top left of the screen.
9 E-mail the saved report to your Agilent Service Representative for evaluation,
and for validation of your personal computer for Agilent Software Systems
Installs.
NOTE
If the firewall is controlled by security software, the Site Preparation Tool cannot
read the firewall settings because of security limitations and will display Status
"Fail" for the firewall settings.
In this case, make sure the firewall is disabled and enter the status in the Site
Preparation Tool report manually.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 11
1 Prepare your PC
Install and Configure Third Party Tools
Install and Configure Third Party Tools
Certain third party tools must be installed and configured on your PC. Some of
these tools can be installed directly from the Installation screen if you run the
OpenLab CDS ChemStation Edition Installer.
Install and Configure Adobe Acrobat Reader
Check if an existing Acrobat Reader version must be uninstalled
1 If Acrobat Reader is already installed, check its version number.
a Open Acrobat Reader.
b Select Help> About Adobe Acrobat Reader.
2 The version number has several ranges. The Continuous or Classic track is
indicated by the first two digits of the third range.
20 = Continuous track
30 = Classic track
NOTE
If an older version of Acrobat Reader (2016 or lower, XI or lower) or a Reader
version from the Continuous track (even with Acrobat Reader 2017) is installed
on your system, you must uninstall it first. Otherwise, the installation below
would result in an Acrobat Reader version that pushes automatic updates.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 12
1 Prepare your PC
Install and Configure Third Party Tools
Install Adobe Acrobat Reader 2017 (Classic Track):
1 Under Third Party Tools, select Adobe PDF Reader.
2 The Acrobat Reader setup screen appears. Click Install to continue.
3 If Acrobat Reader was successfully installed, click Finish to exist the setup
screen.
Alternatively, you can install Acrobat Reader from the ChemStation
installation medium. It is available under Disk1/Tools/Adobe Reader.
Run AcroRdr_MUI.bat and follow the instructions of the Adobe Reader Setup
wizard.
4 Open the newly installed Adobe Acrobat Reader to confirm the Adobe Reader
license agreement. You will be asked only this one time.
Configure Adobe Acrobat Reader in Windows:
1 Set Adobe Acrobat Reader as your default PDF viewer.
a Windows 7: In the Microsoft Control Panel, navigate to Default Programs>
Set default programs, select Adobe Acrobat Reader from the list, and click
Set this program as default.
OR
Windows 10: Navigate to Settings> Apps> Default Apps, select Choose
default apps by file type. In the list, navigate to .pdf and select Adobe
Acrobat Reader as default.
b Select Adobe Acrobat Reader as default PDF viewer.
2 Disable the Adobe Update Service.
a In the Windows Start menu, search for Services. In the results, click the
Services Desktop app.
b Stop the Adobe Acrobat Update Service.
c In the service properties, set the startup type to Manual.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 13
1 Prepare your PC
Install and Configure Third Party Tools
Ensure correct Adobe Acrobat Reader settings:
1 Click Edit> Preferences in Acrobat Reader.
2 Select the General category.
a Ensure that the Open documents as new tabs in the same window check box
is cleared. If it is selected, Acrobat Reader may interfere with the
ChemStation Report Viewer feature.
b Clear the Show me messages when I launch Adobe Acrobat Reader check
box. If it is selected, Adobe messages may interfere with the ChemStation
software.
3 Select the Security (Enhanced) category.
a Ensure that the Enable Protected Mode at startup check box is selected.
b Ensure that the Enable Enhanced Security check box is selected.
4 Select the Tracker category.
Ensure that the Show notification icon in system tray check box is cleared.
Update Acrobat Reader regularly
1 Update Adobe Acrobat Reader on a regular basis to avoid push notifications
from Adobe.
We recommend to include Acrobat Reader updates in a regular update
schedule. Acrobat Reader updates usually come on a quarterly basis.
2 To update, open Acrobat Reader, and click Help> Check for Updates....
3 After an update, ensure that the Adobe Acrobat Update Service is still stopped.
Install Keysight IO Libraries Suite
IO Libraries are required only for LC/MS and CE/MS instruments.
To install IO Libraries:
1 From the installer, select Installation.
2 Under Third Party Tools, select Keysight IO libraries.The Keysight IO libraries
setup screen opens.
3 Click Install to continue.
4 When Keysight IO libraries have been successfully installed, click Finish to exit
the setup screen.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 14
1 Prepare your PC
Install and Configure Third Party Tools
Install the .NET Framework
Install .NET 3.5
If .NET 3.5 is not installed on your system, its installation will automatically be
triggered by the installation wizard. Follow the procedure below to install it in
advance.
1 Go to the Microsoft Control Panel.
In the Windows Start menu, enter "Control Panel" in the Search programs and
files field (alternatively, press [Win+R] and enter "Control panel"). To view all
items in the Control Panel view, select Small icons in the View by field.
2 Go to Programs and Features.
3 Go to Turn Windows features on or off.
4 Enable .NET 3.5 as follows:
This requires an internet connection.
For Windows 10: Expand the .NET Framework 3.5 (includes .NET 2.0 and
3.0) node,
For Windows 7: Expand the Microsoft .NET Framework 3.5.1 node,
in both cases select the Windows Communication Foundation Non-HTTP
Activation check box.
Figure 1 Enable .NET 3.5 (Win 10)
NOTE
If this does not work as expected, or the computer has no internet access, install
.NET 3.5 from the Windows installation media. For Windows 10, see details 10
under https://support.microsoft.com/en-us/kb/2734782.
If you do not have installation media, create them as described under
https://www.microsoft.com/en-us/software-download/windows10.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 15
1 Prepare your PC
Install and Configure Third Party Tools
Install .NET 4.7
If .NET 4.7.2 is not installed on your system, its installation will automatically be
triggered by the installation wizard. When manually installing .NET 4.7.2,
Windows requires write access to the installation files. Direct installation from
USB media is therefore not possible.
1 Copy the folder Disk1\Tools\DotNet4.7 to a local disk.
2 Run dotNetFx_Full_x86_x64.bat.
3 Follow the installation wizard.
4 Restart the computer.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 16
1 Prepare your PC
Configure a Printer
Configure a Printer
Physical Default Printer
Make sure that a default printer is configured in Windows. Every Windows user
who runs ChemStation needs a default printer configured in the user profile. The
printer driver must be for a physical printer, even if the printer is not connected.
Configuring a to-file printer such as a PDF or XPS writer is not sufficient.
This is done via the Microsoft Control Panel. If no default printer is configured the
following problems may occur:
Printing of a report preview will fail
the Copy To Clipboard menu will have an error
the Custom Report View will have problems with new templates
PDF XChange printer
During the installation of ChemStation C.01.10, a PDF XChange 6 printer driver is
installed (revision 6.0.317.1).
Print Limitations
The maximum number of pages for one print job is 1500; for example, a
sequence summary report is one print job.
When printing Multi-Page chromatograms (specified in the Specify Reports dialog
box), the maximum number of pages that can be printed properly depends on the
resolution of the printer. 300 dpi allows ten pages per chromatogram, 600 dpi
allows up to six pages per chromatogram.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 17
1 Prepare your PC
Configure a Printer
Add a Network Printer as a Local Printer
Use physical printers to ensure correct function. Follow the steps below to add a
network printer as a local printer. In Secure Workstation for OpenLab CDS
ChemStation Edition, you will be able to choose this printer as a default printer
when configuring instruments.
1 In the Microsoft Control Panel, navigate to Devices and Printers.
2 Click Add a Printer.
3 Click The printer I want isn't listed, then select Add a local printer and click Next.
4 Select Create a new port, then select Local Port for the port type and click Next.
5 For Port Name, enter the network path to the printer. The network path
consists of two slashes, the computer name or local IP address of the PC
sharing the printer, and the share name of the printer. For example, \\
PTPRINT\PG5-B or \\192.168.1.100\hpprinter.
6 Select the appropriate printer driver and click Next. If the exact model is not
listed, try the closest model number or a generic printer.
7 Follow the rest of the wizard.
NOTE
The print quality of graphics and pictures depends on the printer hardware,
maintenance state, configuration, printing options, and paper quality. Resizing
pictures and graphics during the print process may lead to a reduced print
quality. Choose paper format and page margins that are suitable for your printer.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 18
1 Prepare your PC
Before You Begin the Installation of ChemStation
Before You Begin the Installation of ChemStation
To simplify installation of the software, it is helpful to decide on some
configuration options before you begin the actual software installation.
1 Decide on a computer name.
The computer name will be reflected in the instrument configuration. To avoid
considerable effort, it is recommend to keep the computer name unchanged
after installing Secure Workstation for OpenLab CDS ChemStation Edition.
2 Make sure that you use only the following characters for user names:
3 May sure you have administrator privileges for your workstation. Power user
privileges are not sufficient (the installation does not start).
4 Decide on a directory location to store all files related to the data system
software, including data, methods, sequences, and configurations.
5 Decide on the software delivery approach you want to use:
Install directly from the USB medium — Insert the USB medium directly at
the workstation computer. (Recommended)
Copy installation files to a centralized location — You can use the utility to
copy the installation files, for example, to a network share folder, and run
the installation from that location. However, some networks may interfere
with installation (see “Copy installation Files to a Centralized Folder for
Installationon page 43).
NOTE
To make sure that a DNS server can resolve the computer name, follow the
internet standard for protocols(RFC952) and use only the following characters:
Letters (a-z, A-Z)
Digits (0-9)
Hyphen (-)
Do not use an underscore.
A-Z, a-z, 0-9, _ (underscore), - (hyphen)
Secure Workstation for OpenLab CDS ChemStation Edition Installation 19
1 Prepare your PC
Before You Begin the Installation of ChemStation
6 Before you install the software, check that your computer meets all
requirements. You can access the OpenLab CDS ChemStation Edition
Requirements guide (CDS_CS_HW-SW-Requirements.pdf) from the
Documentation and Learning platform.
a On the installation media, go to Disk1\DOCS\ and open welcome.html to
access Documentation and Learning.
b Select your language.
c On the Welcome page, navigate to Site Preparation> Site Preparation and
Requirements, and select CDS ChemStation Edition Hardware and Software
Requirements.
Use this PDF to check that your settings comply with the network
requirements, and to determine whether your hardware and software will
support the system.
7 Install all required hardware, including any A/D connections, interfaces,
instrument detectors, and communication cables.
8 To allow the successful installation and activation of all components, make
sure that the LAN-cable is connected and a network is available. A local
network is sufficient. If you install the Secure Workstation without a network
cable, the Content Management activation will fail, and trying to access
Content Management will return an error.
9 Make sure that the antivirus software is disabled during the installation.
10 Ensure that the latest graphic card driver is installed. Install the latest
vendor-specific driver. Do not use any generic driver.
11 Disable hardware acceleration. Ensure that the following registry key is
present and correctly set:
[HKEY_CURRENT_USER\SOFTWARE\Microsoft\Avalon.Graphics]
"DisableHWAcceleration"=dword:00000001
12 Make sure .NET 3.5 and .NET 4.7 are activated as Windows features.
For installation instructions, see “Install .NET 3.5” on page 14 and “Install .NET
4.7” on page 15.
13 Check the Agilent Service Notes for software updates for your installation
package and software products. Service Notes are available from your Agilent
support representative.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 20
1 Prepare your PC
Configure the Operating System
Configure the Operating System
PC Configuration Checker
The OpenLab CDS ChemStation Edition configuration check tool helps to prepare
or troubleshoot the Operating System configuration and to prevent computer
problems.
This tool offers two types of configuration checks:
Mandatory: It checks and repairs all mandatory settings which should be
applied before installing OpenLab ChemStation Edition.
Optional: It checks the settings for performance and usability. The
configuration settings are user-specific and must be set separately for each
user.
The tool comes as .diagcab file, which is a file format used with the Microsoft
Windows Troubleshooting Platform (WTP) program. The Microsoft Windows
Troubleshooting Platform (WTP) is a platform to locate and fix hardware and
software settings in Windows. It is used specifically for diagnosing and repairing
computer settings.
In general, .diagcab files are useful for deploying troubleshooting packs because
they are self-contained and require no installation. The .diagcab file name
extension is a registered file name extension that can be executed by WTP.
Depending on whether it is a configuration check for mandatory or optional
settings, different .diagcab files must be executed:
To perform the configuration check for mandatory settings, you must use the
file Agilent.Wtp.ChemStation.WindowsConfiguration.diagcab. This file is
located in Disk1\Tools\Windows Configuration Tools\Mandatory Settings.
To perform the configuration check for optional settings, you must use the file
Agilent.Wtp.ChemStation.WindowsConfiguration.diagcab. This file is located in
Disk1\Tools\Windows Configuration Tools\Optional Settings.
NOTE
If User Account Control (UAC) is switched on, some configuration steps will
require active confirmation to continue.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 21
1 Prepare your PC
Configure the Operating System
1 Before starting the configuration check, copy the .diagcab files to a local disk.
2 Run the tool as administrator to ensure all settings will be applied: On the start
page, select Advanced, then click Run as administrator.
After the settings have been applied, you can generate a report to see which
settings have been changed.
3 To create the report:
a On the Troubleshooting Complete page, click View details.
b Click on the print button to generate the report.
4 Reboot the PC after running the configuration tool.
Manual Configuration Steps
Some Windows changes within this document are mandatory for Secure
Workstation for OpenLab CDS ChemStation Edition to work properly on a
Windows system. Some changes will optimize application performance. Other
changes will have an impact on usability.
The configuration settings are grouped according to their relevance for the
Windows system in the Mandatory, Performance, and Usability categories.
Depending on your operating system, some special configuration steps may be
required.
The following sections describe the configuration steps for each category
separately. Configure your Windows system accordingly. Alternatively, run the
OpenLab CDS configuration check tool (see “PC Configuration Checker” on
page 20).
NOTE
While using the Configuration Checker:
Ensure that this computer is not turned off by another user.
Win 7: Ensure that the menu bar is enabled (click Organize> Layout, and select
Menu bar).
It is very important that you reboot the PC after running the configuration tool.
NOTE
The following Windows 10 descriptions apply to Windows 10 Build 1809. The
settings for other builds may differ slightly.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 22
1 Prepare your PC
Configure the Operating System
Mandatory Configuration Steps
The procedures of this section contain all the configuration steps required for
installation and necessary to ensure proper installation and avoid data loss.
In the Microsoft Control Panel:
In the Windows Start menu, enter "Control Panel" in the Search programs and files
field (alternatively, press [Win+R] and enter "Control panel"). To view all items in
the Control Panel view, select Small icons in the View by field.
1 Administrative Tools: Configure security options:
a Double-click Local Security Policy.
b Navigate to Security Settings> Local Policies> Security Options.
c Double-click the following policy listed in the right hand panel: Network
Access: Sharing and security model for local accounts.
d In the displayed dialog select the following item from the drop-down list:
Classic - local users authenticate as themselves.
2 Network and Sharing Center:
a Select Change adapter settings. Right-click your Ethernet Adapter, then
select Properties. On the General tab, click Configure.
b On the Power Management tab, clear all check boxes.
3 Power Options:
a As preferred plan select High performance.
b Click Change Plan settings.
c Set the option Put the computer to sleep to Never.
d Click Change advanced power settings.
e Open the nodes for Hard disk> Turn off hard disk after.
f Set the Minutes to 0 (=Never).
NOTE
For Windows 7 users:
Ensure that Windows hotfix KB2999226 (Update for Universal C Runtime in
Windows) is installed on your system before installing ChemStation. See
https://support.microsoft.com/en-us/help/2999226/update-for-universal-c-ru
ntime-in-windows.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 23
1 Prepare your PC
Configure the Operating System
4 Programs and Features:
a Click Turn Windows features on or off.
b Select the Telnet Client check box.
c Select the TFTP Client check box.
d Reboot the PC if necessary.
5 Windows 7 only:
Administrative Tools: Disable Services.
Double-click Services.
For the following services, set the startup type to Disabled:
Application Experience
Desktop Window Manager Session Manager
Other Windows settings:
1 To configure Windows logon options, right-click Start, select Run from the
context menu, then type gpedit.msc in the Run field.
a Navigate to Local Computer Policy> Computer Configuration> Administrative
Templates> System> Logon.
b Set Hide entry points for Fast User Switching and Always use classic logon to
Enabled.
2 Windows 10 only: Start> Settings> Update & Security> Windows Security:
a Under App & browser control:
Turn off SmartScreen under Check apps and files, SmartScreen for
Microsoft Edge, and SmartScreen for Windows Store apps.
b Under Virus & threat protection:
Ensure that a protection tool is turned on.
If you want to use Windows Defender: Enable and configure it in the
Windows Control Panel under Windows Defender Firewall> Advanced
Settings.
3 Windows 10 only: Start> Settings> Devices> Printers & scanners: Choose
default printer.
a Turn off Let Windows manage my default printer.
b Click the printer that you want to set as default, then click Manage.
c In the printer settings, click Set as default.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 24
1 Prepare your PC
Configure the Operating System
4 Windows 10 version 1903 only:
If the Net.TCP Port Sharing Service is disabled, enable it by running the
PowerShell command:
Enable-WindowsOptionalFeature -online -All -FeatureName
WCF-TCP-Activation45
Performance Configuration Steps
The procedures of this section contain configuration steps that improve system
performance. These settings are optional.
In the Microsoft Control Panel:
1 System: Change system properties:
a Click Advanced system settings.
b On the Advanced tab under Performance, click Settings.
Advanced tab > Virtual Memory: For optimum performance use the
Change button to adjust the paging file size to a value of 2 to 3 times of
the physical RAM on the PC. If possible locate the paging file on a drive
different from the system installation drive.
Data Execution Prevention tab: Select Turn on DEP for essential Windows
programs and services only.
Click OK to close the Performance Options dialog.
c For Windows 10 and Windows 7 only: On the System Protection tab:
Make sure that Protection is turned off. If required, click Configure and
select Disable system protection.
2 Windows 10 and Windows 7 only: Indexing Options: Disable indexing.
Click the Modify button. Select Show all locations, and clear all drives and
locations.
Other Windows settings:
1 Windows 10 only: Start> Settings> Personalization> Colors: Turn Transparency
effects off.
Usability Configuration Steps
The procedures of this section contain configuration steps that improve the
usability of the Windows application. These settings are optional.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 25
1 Prepare your PC
Configure the Operating System
In the Microsoft Control Panel:
1 File Explorer Options (Folder Options for Windows 7): In the View tab:
Select Always show menus.
Select Display the full path in the title bar.
Clear Hide extensions for known file types.
Clear Use Sharing Wizard.
2 System: Change system properties:
a Click Advanced system settings.
b On the Advanced tab under Startup and Recovery, click Settings.
In the System startup section:
Change both Time to display ... fields from 30 to 3 sec.
For Windows 10 and Windows 7 only: In the System failure section:
a Select Automatically restart.
b In the Write debugging information subsection, select Kernel memory
dump from the drop-down list.
Click OK to close the Startup and Recovery dialog.
c On the Advanced tab under Performance, click Settings. On the Visual
Effects tab:
Select Adjust for best performance, then click Apply.
Select Custom, then select the following check boxes for better
usability:
Show shadows under mouse pointer
Show shadows under windows
Smooth edges of screen fonts
Click OK to close the Performance Options dialog.
Other Windows settings:
1 Enable the navigation pane:
Windows 10: Open Windows Explorer, then select View> Navigation pane
from the ribbon and make sure that Navigation pane is selected.
Windows 7: Open Windows Explorer, then select Organize> Details> Layout
and make sure that Navigation pane is selected.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 26
1 Prepare your PC
Configure the Operating System
2 Windows 7 only: Configure general layout of the Start Menu: (right-click Start>
Properties)
a Start Menu Tab: In the Privacy section select both items
b Start Menu Tab > Customize button: In Customize Start Menu dialog:
Clear the following option:
Favorites menu
Select the following options:
Computer Display as a link
Connect To
Control Panel: Display as a menu
Default Programs
Devices and Printers
Documents: Display as a link
Enable context menus and dragging and dropping
Games: Don't display this item
Help
Highlight newly installed programs
Music: Don't display this item
Network
Open submenus when I pause on them with the mouse pointer
Personal folder: Display as a link
Pictures: Display as a link
Run command
Search other files and libraries Search with public folders
Search programs and Control Panel
Sort All Programs menu by name
System administrative tools: Display on the All Programs menu and in
the Start menu
Use large icons
Secure Workstation for OpenLab CDS ChemStation Edition Installation 27
1 Prepare your PC
Configure the Operating System
3 Windows 10 and Windows 7 only: Configure Windows logon options:
Right-click Start, select Run from the context menu, then type gpedit.msc in
the Run field.
a Navigate to Local Computer Policy> Computer Configuration> Administrative
Templates> System> Logon.
b Set Don't display the Getting Started welcome screen at logon to Enabled.
4 Recycle Bin Properties: (right-click on desktop icon Recycle Bin, then select
Properties) Select the following options:
Custom size: Select a size corresponding to approximately 10% of the
complete disk space for the drive.
Select Display delete confirmation dialog.
Repeat these steps for all drives of your computer.
5 Select Internet Explorer as default Web browser:
Windows 10 only:
a Under Start> Settings> Apps> Default Apps, select the Internet Explorer
as default Web browser.
Windows 7 only:
a In the Microsoft Control Panel under Default Programs> Set your default
programs, select Internet Explorer from the list.
b Click Set this program as default.
6 Windows 10 only: Start> Settings> System> Tablet Mode:
a For When I sign in, select Use desktop mode.
b For When this device automatically switches tablet on or off, select Don't ask
me and don't switch.
7 Windows 10 only: Start> Settings> Apps> Offline Maps: Turn Metered
connections and Map updates off.
8 Windows 10 only: Start> Settings> Privacy:
a On the General page, turn off the following:
Let apps use advertising ID to make ads more interesting to you based on
your app usage (turning this off will reset your ID)
Let website provide locally relevant content by assessing my language list
Let Windows track app launches to improve Start and search results
b On the Location page, make sure Location for this device is off. If not, click
Change to turn it off.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 28
1 Prepare your PC
Configure the Operating System
9 Windows 10 only: Start> Settings> Personalization: Disable advertising info:
a On the Lock screen page:
Under Background, select Picture or Slideshow.
Turn off Get fun facts, tips, tricks, and more on your lock screen.
Turn off Show lock screen background picture on the sign-in screen.
b On the Start page:
Turn off Occasionally show suggestions in Start.
10 Windows 10 only: Start> Settings> Personalization: In the Taskbar tab, under
Taskbar buttons select Combine when taskbar is full.
This will simplify switching between open CDS instances.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 29
1 Prepare your PC
Configure the Operating System
Edit the Security Settings for LC/MS and CE/MS
Systems
The following sections summarize all security settings required for LC/MS and
CE/MS. All other security settings are set automatically by the OpenLab CDS
ChemStation Edition Installer.
OpenLab CDS ChemStation Edition needs to be installed using an operating
system Administrator user account. To run the system with the configured
options outlined in this document, all users and power users should use the
same settings.
Firewall
On PCs controlling LC/MS or CE/MS systems, it is recommended that you turn
off the firewall.
Advanced Network Settings
The communication with the MS is sensitive to the order NICs in the Adapters and
Bindings dialog.
1 Go to Start> Control Panel.
Go to Network and Sharing Center
1
.
2 Click Change adapter settings.
3 Press ALT to bring up the menu.
1
View the items by icon to see a list of all items.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 30
1 Prepare your PC
Configure the Operating System
4 Select Advanced Settings....
Figure 2 Advanced Settings dialog, Adapters and Bindings tab
5 In the Adapters and Bindings tab (see Figure 2 on page 30):
a Make sure that the Local Area Connection pertaining to the LC/MS or
CE/MS NIC is the first item in the list of connections
NOTE
The names of your LAN cards may differ from those shown in the example. You
can tell which LAN card is the Instrument LAN by comparing the IP addresses
assigned to the LAN cards.
The LAN pertaining to the company intranet > internet will typically have an IP
address assigned by organizations Static IP policy or by a DHCP server. Check
with your network administrator.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 31
2 Install the Software
Prepare for Installation 32
Manage Passwords 32
Run the OpenLab CDS Installation Wizard 33
License Agreement Screen 33
Installation Folder Screen 33
Installation Type Screen 34
Summary Screen 35
The installation is automated by the OpenLab CDS ChemStation Edition Installer.
This tool installs all components needed.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 32
2 Install the Software
Prepare for Installation
Prepare for Installation
1 Make sure that the antivirus software is disabled during the installation.
2 Do not run the Windows Update Service during installation. Make sure that no
Windows updates are performed during the installation of ChemStation.
3 Make sure that no system reboot is pending.
Pending reboots are indicated both in the Configuration Checker (see “PC
Configuration Checker” on page 20) and in the Site Preparation Tool (see “Run
the Site Preparation Tool” on page 10).
4 To begin installation, navigate to \Disk1\Setup.bat. Right-click the file and run
it as administrator to proceed to the Planning screen.
Manage Passwords
During the installation, you will need to provide passwords for the following roles:
Administration of PostgreSQL database
Administration of Shared Services database
Administration of Content Management database
Ensure that access to these passwords is restricted corresponding to your
company's password management policy.
These passwords are only required for advanced database administration. They
are not required for normal lab operation or lab administration. For normal lab
operation you only need a Control Panel login, which is initially provided by the
software (see “First Loginon page 60).
Secure Workstation for OpenLab CDS ChemStation Edition Installation 33
2 Install the Software
Run the OpenLab CDS Installation Wizard
Run the OpenLab CDS Installation Wizard
License Agreement Screen
1 From the OpenLab CDS ChemStation Edition Installer, select Installation.
2 Select Secure Workstation for OpenLab CDS ChemStation Edition.
3 The OpenLab CDS Installation Wizard opens. Read the terms of the License
Agreement. The installer provides a printable PDF of the license agreement
under the Resources option of the main menu.
4 Select I agree with the terms and conditions. You cannot proceed with
installation unless you agree to these terms.
5 Select Next to proceed to the Installation Folder screen.
Installation Folder Screen
1 Type the folder name or browse to the directory where you want to store the
application components (typically this is in the programs folder). Folders
must have English names.
2 To install e-familiarization documents during installation, select Include
e-familiarization. This will extend the installation time considerably.
3 To run an installation verification as part of this installation, select Run
Software Verification. The Software Verification Tool provides documentary
evidence that your system has been built and installed correctly, and that all
design specifications have been met. You can run the Software Verification
Tool at a later time if you prefer (see “Run an Software Verification after
Software Installationon page 37).
4 Select Next to proceed to the Installation type screens.
NOTE
Installations into the root of a drive may cause problems during operation and
are not supported.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 34
2 Install the Software
Run the OpenLab CDS Installation Wizard
Installation Type Screen
1 Under OpenLab CDS ChemStation Edition, provide the following folder paths:
Installation folder: directory where you want to store the ChemStation
application components. Typically this is in the programs folder. Folder
names must be entered without spaces.
Instrument data folder: Instrument specific data such as methods,
sequences, and results. The default data path is the public documents
folder. If you plan to activate the Secure File I/O feature, this folder must
not be shared at a later point in time. If your PC is equipped with a second
disk drive, it is recommended to change the default data path to this
alternative drive. Using a second disk drive increases the performance.
2 Select Next to proceed to the Additional items screen.
3 Under Content Management Content folder, Content Management Index Folder,
and Content Management Archive Path, type the required folder names or
browse to the directories. Folder names must be entered without spaces.
4 Under PostgreSQL, type the folder name or browse to the directory where you
want to store the PostgreSQL database files. For example, C:\PostgreSQL
Data.
Provide the password that will be set for the PostgreSQL administrator. Note
that the password must be at least eight characters and contain at least one
lower case letter, one upper case letter, one digit, and one special character.
5 Select Next to proceed with the Databases screen.
6 Under Shared Services and OpenLab Content Management, type the database
names or use the default names, and create user and administrator accounts
as required.
7 Select Next to proceed to the Summary screen.
NOTE
Use a drive with sufficient space for the Data Store content.
It is recommended that you use the same drives for the Content Management
Content folder and the Content Management Archive Path, but a different drive for
the Content Management Index Folder. For example:
D:\DSContent
C:\DSIndex
D:\DSArchive
Secure Workstation for OpenLab CDS ChemStation Edition Installation 35
2 Install the Software
Run the OpenLab CDS Installation Wizard
Summary Screen
1 An installation summary is shown. Select Back as necessary to change
installation settings, or Cancel to cancel the installation.
Select Start to begin installation.
2 The system performs an automated system check before it proceeds with the
listed activities.
If a system check passed message appears, installation continues.
If a system check failed message appears, you can either:
Decline to view the system report, and continue installation.
Decline to view the system report and postpone installation.
View the system report, and decide to continue installation.
View the system report and postpone installation until the problem is fixed.
3 If an installation verification was completed as part of this installation, review
the Software Verification Report. If the report indicates failure, verify the
computer requirements and reinstall the data system. Do not use the system
until the Software Verification Report gives a ‘pass’ result.
4 Click Next to proceed to the Installed Features screen.
5 Click Finish to close the installation wizard.
6 Reboot the computer.
NOTE
To view the system report as PDF file, Adobe PDF Reader must be installed.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 36
3 Post Installation Tasks
Run an Software Verification after Software Installation 37
Allow ChemMain Through Firewall 38
Configure the Antivirus Program 39
This chapter describes tasks that are relevant after finishing the installation.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 37
3 Post Installation Tasks
Run an Software Verification after Software Installation
Run an Software Verification after Software
Installation
The Software Verification Tool (SVT) provides documentary evidence that your
system has been built and installed correctly, and that all design specifications
have been met. You do not need to run the software verification again if it has run
successfully at the end of the installation.
1 Using your Windows operating system, go to Start> All Programs> Agilent
Technologies> Software Verification Tool.
2 Select the components to qualify.
3 Select Qualify.
The system will run the application and generate a Software Verification
Report.
4 If the report indicates failure, verify the computer requirements and reinstall
the data system.
Do not use the system until the Software Verification Report gives a ‘pass’
result.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 38
3 Post Installation Tasks
Allow ChemMain Through Firewall
Allow ChemMain Through Firewall
Carry out the following procedure to prevent ChemStation from functioning
incorrectly due to firewall restrictions. Consider using a second network card to
isolate the instrument's data traffic, and carry out the following procedure for that
second network card only.
Alternatively, ensure that all required firewall ports are open. For details, refer to
the OpenLab CDS ChemStation Edition Requirements guide
(CDS_CS_HW-SW-Requirements.pdf).
1 In the Microsoft Control Panel, click Windows Defender Firewall
1
.
2 Click Allow an app or feature through Windows Defender Firewall.
3 Click Change settings.
The Allow another app... button becomes active.
4 Click Allow another app....
5 Click Browse... and navigate to C:\Program Files (x86)\Agilent Technologies\
ChemStation\CORE.
6 Select ChemMain.exe and click Open.
7 Click Add.
8 In the list of allowed apps and features, find ChemStation ChemMain, and
select the check boxes for all three network types.
9 Confirm your changes.
10 Repeat the procedure for ChemMainAcq.exe.
1
View the items by icon to see a list of all items
Secure Workstation for OpenLab CDS ChemStation Edition Installation 39
3 Post Installation Tasks
Configure the Antivirus Program
Configure the Antivirus Program
Be sure to open the firewall ports listed in the Firewall Settings in the OpenLab
CDS ChemStation Edition Requirements guide.
In order for the OpenLab software to function correctly, you should configure any
antivirus real time protection software with the following folder exclusions. They
should only be scanned while the instruments are idle and no data acquisition
takes place. Refer to your specific antivirus software documentation on how to
configure folder exclusions.
NOTE
Running antivirus programs might influence the behavior and performance of
your computer. Some virus scanners might cause issues when used with
ChemStation. The application is tested with Symantec Endpoint Protection 14.0
MP2 and with Microsoft Security Essentials.
Process Directory File name
Data acquisition %public%\Documents\Chem-
Station
(or the corresponding folder for
instrument data that you pro-
vided during installation)
Data, methods, sequences,
reports etc.
Standard reports %temp% for Windows users
(=Users' temp directory)
~p3d*.tmp
~job*.tmp
Hpspl00.que
CDS intelligent reports %LOCALAPPDATA%
%APPDATA%
%PROGRAMDATA%
Files on:
Agilent
Agilent Technologies
Agilent_Technologies,_Inc
IsolatedStorage
Temp
e.g.:
C:\Users\xxxxx\
AppData\Local\ Agilent
Technologies\Intelligent
Reporting\RawDataFile-
Cache
Secure Workstation for OpenLab CDS ChemStation Edition Installation 40
3 Post Installation Tasks
Configure the Antivirus Program
If your antivirus software includes program or executable deny execution
settings, ensure that the following program files are not denied execution. You
can use the windows search feature to find the specific folder each program file
is located in.
agilentiolibrariesservice.exe
chemmain.exe
chemmainacq.exe
apg_top.exe
iprocsvr.exe
iproc8491.exe
msinsctl.exe
httpdmsd.exe
epcsetup.exe
AcroRd32.exe
Acrobat.exe
NOTE
Depending on your specific configuration, some of the listed folders or files may
not exist on your system.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 41
4 Optional Procedures
Install Additional Software and Drivers 42
Copy installation Files to a Centralized Folder for Installation 43
Add Shortcut to Public Folder 44
This chapter contains information on the Additional Drivers and Software wizard
and other helpful procedures.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 42
4 Optional Procedures
Install Additional Software and Drivers
Install Additional Software and Drivers
OpenLab CDS ChemStation Edition offers a wizard to help you installing
additional software, such as the ADFExport Plug-in, or drivers for third-party
instruments. To open the wizard, go to Start> All programs> Agilent Technologies>
OpenLab Additional Software and Drivers. Follow the wizard to install the required
software.
Prepare Network Drives
If the additional software is located on a network drive, you must prepare the
network drive to make it accessible by the wizard. Without this preparation,
Windows security prevents the wizard from accessing those drives.
1 Map the drive to a letter.
For example, map the drive as Z: using the shared path "\\<machine-name>\
OpenLabCDS".
This maps the drive for the logged-in user.
2 Open the command prompt in elevated mode (run as administrator), and map
the drive using the net use command.
For example,
net use Z: "\\<machine-name>\OpenLabCDS"
This maps the drive for the local administrator account. The mapped drive is
now visible to both logged-in user and administrator, and can be selected in
the wizard.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 43
4 Optional Procedures
Copy installation Files to a Centralized Folder for Installation
Copy installation Files to a Centralized Folder for
Installation
Completing this step will enable you to run an installation from a network share.
1 From the installer Planning screen, select Installation from the sidebar menu.
2 Select Preparation of an Installation from Network Share.
3 At the Network Share screen, browse to a directory and create a destination
folder as follows:
a Select the button with the three dots.
b Navigate to the directory where you want to create the folder.
c Select Make New Folder.
d Type in the folder name.
e Select OK. The system will return you to the Network Share screen, with the
path displayed.
f Select the Include Secure Workstation Installation media check box.
g Select Start.
4 When processing is complete, copy the files to the local drive or map the
location to a network drive.
5 Close the application and navigate to the directory and folder you created.
Open the folder.
6 Select the Disk 1 folder, then execute Setup.bat to run the application.
The system will display the installer Planning screen.
NOTE
Installations into the root of a drive may cause problems during operations and
are not supported.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 44
4 Optional Procedures
Add Shortcut to Public Folder
Add Shortcut to Public Folder
By default, user data such as master methods, sequence templates, report
templates, raw data etc. is located in the public documents folder C:\Users\
Public\Documents\ChemStation. You define this folder during the installation.
The system creates a shortcut to the defined folder under Instrument Data in the
Windows Start menu.
In Windows 7, a shortcut to public documents is automatically available in the
Windows Explorer.
In Windows 10, the Windows Explorer is organized differently. To provide easy
access to that folder, we recommend that you pin the Instrument Data folder to
the Start menu.
Figure 3 Windows Start menu in Windows 10
1 Navigate to the Instrument Data shortcut in the Start menu.
2 Right-click the icon, then select Pin to Start.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 45
5 Secure File I/O
About Secure File I/O 46
Prepare Secure File I/O in Windows 47
Enable Secure File I/O in ChemStation 48
Check privileges for ChemStation users 49
Important Notes for Working with Secure File I/O 50
Limitations 51
This chapter describes how to protect your ChemStation folders with Secure File
I/O.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 46
5 Secure File I/O
About Secure File I/O
About Secure File I/O
ChemStation metadata such as validated master methods or sequences
templates must be stored in secured locations in your content management
system. For usage with your instruments these methods or sequence templates
are also stored in various local folders. By default, the content of these folders is
not protected against modification or deletion from local file browsers or from
file dialogs within ChemStation.
To ensure data integrity, enable the Secure File I/O function for ChemStation. All
relevant local paths will then be protected. The relevant paths are set by default,
but can also be configured. The content of protected paths can no longer be
modified by Windows file operations because access is denied for members of
the Windows user group Interactive
1
. The protected folders can only be used
inside ChemStation. If user authentication is enabled, only users with sufficient
privileges can delete the locally stored data for housekeeping purposes. If those
methods or sequences have been actively used with your instrument, copies are
also stored along with results and their associated audit trails.
1
In Windows, all users who log on interactively are automatically assigned to the
Interactive group. The membership persists for the time the users remain logged in.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 47
5 Secure File I/O
Prepare Secure File I/O in Windows
Prepare Secure File I/O in Windows
1 Ensure that all relevant ChemStation folders are on an NTFS file system.
2 Do not use ChemStation folders on removable storage devices (for example,
external disks or USB media).
3 Create a standard user account (local or domain) with the same or fewer
privileges than a normal ChemStation user. Do not grant administration or
power user privileges. Make sure that this user is not a member of any group,
especially not the Interactive group.
This account will be used by the ChemStation for internal file operations. As
the user is not supposed to log on interactively, do not use the option Change
password at next logon.
NOTE
Consider using a password that never expires. If this is not possible due to your
security policy, plan a downtime for ChemStation while updating the password.
As long as the password for this user account is expired, do not use the
ChemStation. ChemStation will not work properly until you have updated the
password in the ChemStation Administration Tool.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 48
5 Secure File I/O
Enable Secure File I/O in ChemStation
Enable Secure File I/O in ChemStation
1 Click Start> All Programs> Agilent Technologies> ChemStation Administration
Tool to open the ChemStation Administration Tool.
2 In the ChemStation Administration Tool, select the enable secure file IO check
box.
3 Provide the credentials of the special user account that you have prepared.
4 Click Apply.
5 A dialog lists all directories that are going to be protected. Click OK to confirm.
For all paths that are protected, a ProtectionInfo.xml file is stored in the root
folder. This protection information file contains the time stamp and the
Windows user who enabled protection - that is either the user who enabled
the path from the ChemStation Administration tool, or the user who
configured a new instrument.
The Protection Information file provides proof for a given folder when the last
activation was performed. Any period of time before that date is unspecified.
The file can be used to demonstrate the uninterrupted protection status of a
folder since it contains the date, time and user who last enabled protection.
Enabling or disabling secure file I/O is not logged to the OpenLab Control
Panel Activity Log or the Windows Event Logs (application or Security).
When the folder is unprotected, the ProtectionInfo.xml file is deleted. Note,
deleting this file will not unprotect the folder.
6 For administrative tasks, you can temporarily disable the protection of
specific folders:
a Shut down all ChemStation sessions.
b In the ChemStation Administration Tool, click Manage Secure File System.
c In the list of directories, find the relevant path and click Unprotect.
The protection information file is deleted to document the interruption of
the protection state.
d After finishing your tasks, click Protect All to restore protection for all
relevant folders.
New protection information files are created.
e Start ChemStation again.
ChemStation will only start if all folders are protected.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 49
5 Secure File I/O
Check privileges for ChemStation users
Check privileges for ChemStation users
In the OpenLab Control Panel, specific privileges can be configured. Several
privileges are relevant for Secure File I/O. They handle occasions when data,
methods, or sequences need to be loaded from unprotected locations (which are
outside of the secure folders). These privileges are enabled in the default
ChemStation Roles to maintain the current workflows.
To ensure data integrity, it is recommended to uncheck these privileges when
Secure File I/O is enabled.
The relevant privileges are:
ChemStation: Data> Load data from not configured path
Required for loading data from unprotected paths, or adding data paths that
are not empty to the ChemStation.
ChemStation: Method> Load method from not configured path
Required for loading a method from an unprotected path, saving a method to
an unprotected path, or adding method paths to the ChemStation
(Preferences dialog).
ChemStation: Sequence> Load sequence template from not configured path
Required for loading sequence templates from unprotected paths, saving
sequence templates to unprotected paths, or adding paths that are to the
ChemStation (Preferences dialog).
Secure Workstation for OpenLab CDS ChemStation Edition Installation 50
5 Secure File I/O
Important Notes for Working with Secure File I/O
Important Notes for Working with Secure File I/O
When using Secure File I/O to protect your ChemStation folders:
Do not share protected folders. Sharing a folder will break the protection.
If your data resides within the public documents folder (default setting), do
not turn on public sharing.
Do not log in interactively with the user account provided in the ChemStation
Administration Tool.
If this user logs in interactively, you will not be able to save any changed data
in ChemStation until the user logs out again.
If the password for this user account has expired, do not use the ChemStation
until you updated the password in the ChemStation Administration Tool.
ChemStation will not work properly as long as the password is outdated.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 51
5 Secure File I/O
Limitations
Limitations
The following features are not supported when Secure File I/O is enabled:
Import/Export eMethods
Migration of non-container data to new result sets
Download files (method, sequence, data) using the content management web
interface, and storing them to a protected folder
These files must be downloaded from within the ChemStation application.
GC Backflush Wizard
G2887BA SIMDIS
M8350AA MatchCompare
M8370AA OpenLab Data Analysis Add-on
Easy SamplePrep
G7818A Cirrus GPC Software
A2Prep software
Method Scouting Wizard
G4218A Evaporative Light Scattering Detector
2D-LC Add-on
Since Secure File I/O protects the folders/directories containing methods,
sequences, data for OpenLab ChemStation, this prevents access from outside
applications as Explorer or Notepad from making edits to ChemStation produced
data, methods, etc. This also includes some of Agilent's Add-on programs such
as A2Prep as these programs need to access ChemStation folders/directories.
Any application that needs to access or write to a protected ChemStation
folder/directory will not be allowed and therefore cannot be used when Secure
File I/O is enabled.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 52
6 Secure Workstation for OpenLab CDS
ChemStation Edition Licensing
About OpenLab Licensing 53
License Types 53
License File 53
Obtain your Software License 54
Generate and download your software license 55
Other Ways to Obtain a License 57
This chapter describes the licensing of a Secure Workstation for OpenLab CDS
ChemStation Edition.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 53
6 Secure Workstation for OpenLab CDS ChemStation Edition Licensing
About OpenLab Licensing
About OpenLab Licensing
License Types
The license file is a collection of Product, Instruments and Add-on licenses (or
activation keys), and is installed to your OpenLab CDS System.
The licenses or activation keys in the license file can either be Shared or Counted:
Shared licenses — system computers and other components can have
shared, or add-on, licenses — because they share a core license.
Counted licenses — these licenses are part of the Secure Workstation for
OpenLab CDS ChemStation Edition floating licensing strategy. They are not
permanently assigned to any one component. Instead they are automatically
assigned to components, such as AICs and instruments, while the
components are starting up. The licenses are automatically returned when
the component is closed. The license management program controls license
issuance and retrieval.
In this case, the only requirement is that a component is licensed while
running. You only need enough licenses for all components running
concurrently, rather than for each installed component.
A startup license for the system allows you to run OpenLab CDS for 60 days after
the installation. In order to run the data system software after the 60-day period,
you must install your license file.
License File
A license file will contain your software license. This file is installed on the
workstation. The license file is bound to this computer, and cannot be moved to
another workstation without regenerating the license in SubscribeNet.
Information in the license file defines the number of instruments and other
options that may be used concurrently with your system.
The most efficient way to manage and maintain your licensing is through the
Internet.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 54
6 Secure Workstation for OpenLab CDS ChemStation Edition Licensing
Obtain your Software License
Obtain your Software License
The most efficient way to and manage and maintain your license is with
SubscribeNet.
Flexera
License
Manager
Secure Workstation for OpenLab CDS ChemStation Edition uses a 3rd party tool
called FlexNet Producer Suite from Flexera to manage the licenses. The required
Licensing server components are installed by default on the Secure Workstation
for OpenLab CDS ChemStation Edition.
License Management in OpenLab Shared Services requires an additional
Windows service to be running. This Windows service is called Agilent OpenLab
License Server. This service must be running on your Secure Workstation for
OpenLab CDS ChemStation Edition.
If you are lacking an Internet connection, see “Other Ways to Obtain a License” on
page 57. Gather the following information from the lavender envelope containing
your Software Entitlement Certificate:
The authorization code label
The URL for SubscribeNet
If you have not received a lavender envelope, contact your vendor or internal
support.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 55
6 Secure Workstation for OpenLab CDS ChemStation Edition Licensing
Obtain your Software License
Generate and download your software license
Use the following procedure to generate and download your license. In case you
do not have internet access, skip to the section “Other Ways to Obtain a
License” on page 57.
If you are a new user who has not registered with SubscribeNet, continue with the
section New Users.
If you have registered with SubscribeNet, skip to the section Users registered with
SubscribeNet.
Prerequisites To generate, download, and install a final license for your product, you will need:
The authorization code label provided in the lavender envelope containing
your Software Entitlement Certificate.
If you have not received a lavender envelope for your product, contact your
vendor or internal support.
The URL for SubscribeNet from the Software Entitlement Certificate.
The host name of the computer where the Control Panel is running.
The MAC address.
To retrieve your MAC address from a computer where Secure Workstation for
OpenLab CDS ChemStation Edition is already installed, open the Control Panel
and browse to the Administration> Licenses section. Use the Copy MAC
Address or Save MAC Address function to obtain the MAC address for license
generation.
During this process you will have to enter the MAC address of your license
server. For workstations, this is the local computer. For client/server systems,
this is the server.
NOTE
If any changes are made to the computer name or domain reference after the
license is installed, remove the license. A new license will need to be created in
SubscribeNet, downloaded, and installed.
NOTE
If the network adapter that provides the MAC address used during license
creation is removed from the machine, your license will no longer be valid. A new
license will need to be generated with a currently available MAC on the license
server.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 56
6 Secure Workstation for OpenLab CDS ChemStation Edition Licensing
Obtain your Software License
New Users
1 Go to https://agilent.subscribenet.com/control/agil/AgilRegisterToAccount
to register the product with SubscribeNet.
2 On the registration page, enter the authorization code from the label and
complete the profile information (required fields are marked with an asterisk
*).
The email address you enter will become your login ID.
3 Click Submit. The system will generate and display an account name for you.
SubscribeNet will send a welcome email with your login ID and password.
4 Log in to SubscribeNet using your login ID and password.
Once you log in, you can use the online user manual link for help with any
questions you have.
5 Select Generate or View licenses from the left navigation bar.
6 Follow the prompts to generate your new license.
You will be prompted for the HOST NAME of the computer. The host name
you enter must match with the network name of the computer where the
Control Panel is running. Do not include any DNS suffix (domain.com)
references in the entered machine name.
7 When the system generated the license, view its details, then click Download
License File. Save the license file to your computer and to a backup location
(such as a portable storage device).
Use your login ID and password when you revisit the Agilent SubscribeNet site
to regenerate a license file, add new authorization codes, or further configure
the license for your system.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 57
6 Secure Workstation for OpenLab CDS ChemStation Edition Licensing
Obtain your Software License
Users registered with SubscribeNet
1 If you already have a SubscribeNet account, use
https://agilent.subscribenet.com/.
Lost your SubscribeNet password? Use
https://agilent.subscribenet.com/control/agil/password to have it emailed
to you.
2 Select the SubscribeNet account associated with this authorization code, if
you have more than one account.
3 From the SubscribeNet navigation pane, select Register Authorization Code.
This will allow you to enter your new authorization code and make available
the new license entitlements.
4 Follow steps 5 through 7 in the previous procedure, New Users, to generate or
view your new licenses.
Other Ways to Obtain a License
If you are unable to generate a license, contact your nearest Agilent technical
support office. A representative will tell you how to submit an OpenLab CDS
License Generation Form in your location.
Offline Licensing
If an internet connection is not available in your laboratory:
You or your local on-site service engineer will collect the necessary information
from you to allow Agilent to create a license account on your behalf. For phone
support in your region, call the sales and service number for your region. See the
Appendix for contact information.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 58
6 Secure Workstation for OpenLab CDS ChemStation Edition Licensing
Obtain your Software License
Required Customer Information for Agilent License Support:
The following information must be provided to Agilent in order to enable us to
create a licensing account on your behalf.
1 Collect Account Information:
Your account name will be your company name and Lab name separated by a
comma. Employee information provided here will be used to define the first
administrator of your account for future access to the system as required.
Please prepare the following pieces of information prior to contacting your
local Agilent sales and service center in order to expedite service:
Company Name
Lab/Department Name
First Name
Last Name
E-mail address
Job Title
Phone #
Address, City, State/Province, Postal Code, Country
2 Collect Authorization Code(s):
The authorization code is an alpha-numeric code provided on a label which is
enclosed in a lavender envelope. If you have received more than one code you
must provide all codes to ensure that all ordered licenses are granted to your
account.
3 Receiving your license:
Once the above information is provided Agilent will then work on your behalf
to generate a license file through SubscribeNet. The license file will either be
sent to your shipping address (on a CD), or your local FSE will deliver it in
person (usually on USB media). Once your license is received follow the below
section on “Install your License” to finish installing your license on your CDS
system(s).
Secure Workstation for OpenLab CDS ChemStation Edition Installation 59
7 Configure Secure Workstation for OpenLab
CDS ChemStation Edition
First Login 60
Add your License to the Control Panel 61
Add Users 62
Add Users to a Role 63
This chapter describes the configuration of Secure Workstation for OpenLab CDS
ChemStation Edition.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 60
7 Configure Secure Workstation for OpenLab CDS ChemStation Edition
First Login
First Login
During the installation, internal authentication is automatically configured, and a
default user 'admin' is created with password ‘openlab'. For security and
compliance reasons, change the password before proceeding any further.
1 Start the OpenLab Control Panel shortcut on the desktop or go to Start> All
Programs> Agilent Technologies> OpenLab> OpenLab Control Panel.
2 Enter your login credentials for OpenLab Control Panel.
a In the Login field enter the user name
admin.
b In the Password field enter the password
openlab.
3 In the Administration view, select My Settings.
4 In the ribbon, click Change My Password.
5 Provide a new password that corresponds to your password policy, and
confirm your settings.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 61
7 Configure Secure Workstation for OpenLab CDS ChemStation Edition
Add your License to the Control Panel
Add your License to the Control Panel
The license must be added to your system using the Control Panel.
1 Start the Control Panel shortcut on the desktop or go to Start> All Programs>
Agilent Technologies> OpenLab Shared Services> Control Panel.
2 Navigate to Administration> Licenses.
3 In the ribbon, click Add License .
4 Choose to install the license by:
Using the license file option to browse to and open the license file (.lic)
saved from the license generation process in SubscribeNet.
Selecting the License Text option and copying the license text from a text
file received into the provided field.
5 Click OK.
The Administration interface in the Control Panel will now display the status of
installed licenses.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 62
7 Configure Secure Workstation for OpenLab CDS ChemStation Edition
Add Users
Add Users
1 From the navigation pane, select Administration> Users.
2 In the Users toolbar, select Create.
3 In the Create User dialog box enter a Name and Description for the user.
4 Select the General tab and enter a Password for the user. Confirm the
password.
5 Under Security policy, set the minimum password length of at least five
characters.
6 Set the expiration period of 180 days or less.
7 Limit the number of allowed login attempts to three.
8 Enter the user's full name, email address, and contact information if desired.
The full name is used in activity log entries and the welcome message at the
lower right of the Control Panel.
9 Select password options:
To prevent the user from changing the password assigned here, enable
the User cannot change password.
To require the user to create a new password the next time they log on,
enable the User must change password at next logon.
To allow the user to use the password assigned here or change it at any
time, disable both the User cannot change password and the User must
change password at next logon.
10 To create a user profile but prevent the user from logging on Control Panel,
enable User disabled.
11 Select OK.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 63
7 Configure Secure Workstation for OpenLab CDS ChemStation Edition
Add Users to a Role
Add Users to a Role
1 From the navigation pane, select Administration> Roles.
2 In the roles window, select the role you want to assign to users.
3 In the roles toolbar, select Edit.
4 In the Edit Role dialog box, select the Members tab.
5 Select Add user or group.
6 In the Search Users and Groups dialog box, select Search to view a list of all
users and groups.
7 Under Search Results, select a user or group and select Add.
8 Select OK.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 64
8 Upgrade Secure Workstation
Planning your Upgrade 65
Convert Classic Instrument Drivers to RC.NET 66
Preparations for the Upgrade 69
License Upgrade 71
Get Upgraded License File 71
Add Upgraded License File to the System 72
Upgrade a Secure Workstation 73
This chapter describes the upgrade of Secure Workstation for OpenLab CDS
ChemStation Edition.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 65
8 Upgrade Secure Workstation
Planning your Upgrade
Planning your Upgrade
A direct upgrade, using the Upgrade Wizard, is supported from ChemStation
C.01.07 SR3 or higher. Older ChemStation revisions must be upgraded to C.01.07
SR3 in a first step. For details on the upgrade to C.01.07 SR3 or SR4, please refer
to the C.01.07 SR3 installation guide.
Note that C.01.10 is supported on Windows 7 SP1 64 bit or Windows 10 only. An
in-place upgrade from Windows 7 or 8.1 to Windows 10 on an existing
ChemStation Workstation is not supported. Before upgrading a Secure
Workstation, make sure that the Windows configuration meets all requirements.
See “Configure the Operating Systemon page 20.
For information on restoring your old database into a new installation, see the
disaster recovery procedures in the Secure Workstation for OpenLab CDS
ChemStation Edition Maintenance Guide (CDS_CS-SecureWS_maintenance.pdf).
Classic instrument drivers, with exception of MSD instrument drivers, are no
longer supported. It is recommended that you convert the corresponding
instrument methods to RC.Net driver methods prior to the upgrade (see “Convert
Classic Instrument Drivers to RC.NET” on page 66).
If your system is configured to use one of the following ELSD drivers (G7102A,
G4261A/B or G4260A/B), refer to the current ELSD Release Note for upgrade
instructions. The version current at the release of ChemStation C.01.10is
available in the Documentation and Learning platform.
The functions related to ADF Export will be available as an Add-On for
ChemStation starting from revision C.01.10. To use the latest version, install the
ADFExport Add-On separately (see “Install Additional Software and Drivers” on
page 42). ADF Export functions have been included in ChemStation C.01.09.
Therefore, if you upgrade ChemStation from C.01.09, always install the latest
ADFExport Add-On on ChemStation revision C.01.10 in order to upgrade the
outdated ADF Export functions.
NOTE
As of C.01.08, the M8370AA OpenLab CDS Data Analysis Add-On is no longer
supported in combination with ChemStation.
Also as of C.01.08, OpenLab CDS ChemStation Edition does no longer control
Agilent 5890 GC instruments.
If you use either of them, do not upgrade to C.01.10. Use rev. C.01.07 SR3 or SR4
instead.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 66
8 Upgrade Secure Workstation
Convert Classic Instrument Drivers to RC.NET
Convert Classic Instrument Drivers to RC.NET
With C.01.10, only RC.NET drivers are available (with exception of MSD
instrument drivers). Instruments using the classic driver must be reconfigured to
use the RC.net driver. Agilent recommends to do this before upgrading to
C.01.10.
To determine which instruments are using the classic driver, go to the OpenLab
Control Panel and select the instrument. Select Configure Instrument in the
ribbon. The instrument is using the classic driver if the Use classic drivers check
box is selected. Perform the following steps to reconfigure the instrument to use
RC.NET.
1 Record the IP address for each classic driver instrument.
2 Verify there is a backup of the methods and data to another location.
3 For your reference: Print the classic driver method settings or save the
method listing to disk.
4 To reconfigure the instrument, select the instrument in the OpenLab Control
Panel.
5 In the ribbon, click Configure Instrument.
6 In the configuration dialog, clear the Use classic drivers check box.
The instrument is moved from the Selected Modules panel to the Configurable
Modules panel.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 67
8 Upgrade Secure Workstation
Convert Classic Instrument Drivers to RC.NET
7 Select the instrument in the Configurable Modules panel, then click the arrow
to add the instrument to the Selected Modules panel again.
8 Double-click the instrument under Selected Modules, and configure the
previously recorded IP address. Click Get Serial Number and Firmware to get
the corresponding entries.
Figure 4 Example for 35900 configuration
9 Launch the newly configured instrument.
NOTE
If a current RC.NET driver is not yet available, you must install it manually
before upgrading OpenLab CDS ChemStation.
For example, follow these steps to install the 35900E ADC RC.NET driver:
a Run the installer.
b Go to Installation and open OpenLab Additional Software and Drivers.
c When asked for the Add-on software, browse to Disk3 of the installation
media, and to the Agilent 35900E RCNet folder to find the Agilent OpenLab
CDS ChemStation 35900 AtoD Drivers.msi file. The corresponding
software will then be listed in the installer.
d Select the software in the list, and continue to install. The installation
verification will automatically follow and should complete without errors.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 68
8 Upgrade Secure Workstation
Convert Classic Instrument Drivers to RC.NET
10 To convert a method to RC.NET driver, load the method. If a dialog like the
following opens, click OK.
Review the method, and if no updates are needed, add a comment such as
"Updated to RC.NET" to the method when saving it.
Methods are converted to RC.NET when loaded. After saving to disk,
converted methods have an additional Agilent folder and
RapidControl.InstrumentConfig file.
Only with RC.NET driver methods
Secure Workstation for OpenLab CDS ChemStation Edition Installation 69
8 Upgrade Secure Workstation
Preparations for the Upgrade
Preparations for the Upgrade
1 Perform a backup of your system by following the procedure described in the
maintenance guide (SecureWS_maintenance.pdf).
2 If not already present from a previous installation: Install .Net Framework
4.7.2 (see “Install .NET 4.7” on page 15).
3 Disable Windows Updates
a Go to Control Panel> Windows Update.
b Click Change settings. In the Important updates section, select Never check
for updates. Clear the other update options.
4 Verify that the locales defined in the postgreSQL database match the locales
defined in the operating system.
a Open the postgreSQL administration tool (pgAdmin)
Start> Postgres 9.x> pgAdminIII
b Right-click Postgres SQL 9.x, select Connect, and type the password for the
postgres user.
NOTE
This setting is important to avoid a system reboot during the upgrade process
NOTE
To perform this step you will need the password for the postgres user that has
been defined during the original installation. It should be part of your personal
documentation.
Contact your local Agilent support representative if you can’t retrieve the
password of the postgres user.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 70
8 Upgrade Secure Workstation
Preparations for the Upgrade
c Expand the Databases list by clicking on (+).
d Click on the different databases. In the Propertiestab, note the language for
Collation and Character Type.
e Open the Windows settings Start> Control Panel> Region and Language>
Format and compare with the language defined in the databases.
f Reboot the computer.
NOTE
Note down the settings in the ChemStation Administration Tool. After the
upgrade, check if the settings still match the original ones before the upgrade.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 71
8 Upgrade Secure Workstation
License Upgrade
License Upgrade
Get Upgraded License File
You will need to upgrade your licenses in SubscribeNet prior to upgrading to the
next version of Secure Workstation for OpenLab CDS ChemStation Edition. We
strongly recommend upgrading your workstation licenses before upgrading the
core software. Standalone workstations which are upgraded to the new core
software version, without a new workstation license, will not work until the new
workstation licenses are added to the OpenLab Control Panel.
If you are under SMA subscription, proceed as follows to upgrade your licenses:
1 During the following process, you will be prompted in SubscribeNet for the
host name or MAC address of the workstation where Secure Workstation for
OpenLab CDS ChemStation Edition is already installed.
To retrieve this hostname and MAC address, open the Control Panel and
browse to the Administration> Licenses section. Note down the host name
and use the Copy MAC Address or Save MAC Address function to obtain the
MAC address.
2 Log into the Agilent Electronic Software and License Delivery
(https://agilent.subscribenet.com/).
3 Navigate to Manage Licenses by Host. In the Host ID field, enter the previously
noted MAC address, and click Search.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 72
8 Upgrade Secure Workstation
License Upgrade
If the relevant host name does not appear, you may be managing your
licenses in multiple SubscribeNet accounts. You will need to log into those
accounts to upgrade those licenses.
4 If your license(s) are eligible for an upgrade, you will see the Upgrade All
button. Otherwise you will need to contact your Agilent Sales Representative
to renew your Software Maintenance Agreement (see “Sales and Support
Assistance” on page 84). To proceed with generating your upgrade license,
click the button.
5 On the Upgrade All Licenses for License Host page, review the data, and
confirm by clicking Upgrade All.
This upgrades the license file to the most current version. SubscribeNet will
send you an email with a new license file.
6 Put the new license file on your system (see Add Upgraded License File to the
Systemon page 72.
If you have multiple standalone Workstations, repeat this step for each
individual workstation.
Note that each workstation's MAC address is the file name. This helps identify
the correct license file to import into the workstation's Control Panel.
Add Upgraded License File to the System
If you have purchased new options, such as additional instrument controls or
client license and regenerated your license in SubscribeNet, the upgraded license
file must be re-applied to the system.
1 Start the Control Panel from any machine connected to the system you want
to install the license for.
2 Navigate to Administration> Licenses.
3 Select a license, then click Remove License .
4 In the ribbon, click Add License .
5 Browse to and open the license file saved from the license generation process
in SubscribeNet.
NOTE
A full restart is required in order for any license to have an immediate effect.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 73
8 Upgrade Secure Workstation
Upgrade a Secure Workstation
Upgrade a Secure Workstation
Prerequisites You are using Secure Workstation for OpenLab CDS ChemStation Edition rev.
C.01.07 SR3 or higher.
Older versions must first be upgraded to C.01.07 SR3. Multiple upgrade steps
may be required.
1 If you protect your folders with Secure File I/O: disable Secure File I/O during
the upgrade.
a Shut down all ChemStation sessions.
b Click Start> All Programs> Agilent Technologies> ChemStation
Administration Tool to open the ChemStation Administration Tool.
c In the ChemStation Administration Tool, clear the enable secure file IO
check box.
2 Run the installer from the same media type (for example, USB or network
share) as you used to install the current version.
3 From the installer Planning screen, switch to the Installation screen.
4 Select Secure Workstation for OpenLab CDS ChemStation Edition.
If a Secure Workstation is already installed, this automatically opens the
upgrade wizard.
5 The workstation license must be upgraded, see “Get Upgraded License
File” on page 71.
Acknowledge that the license has been upgraded and click Next to continue.
6 Select I agree with the terms and conditions. You cannot proceed with the
upgrade unless you agree to these terms. Click Next.
7 In the OpenLab Shared Services Settings for Registration screen, enter the
username and password of a user with system administration privileges.
Click Next.
8 In the Summary screen of the Upgrade Wizard, the components for the
upgrade are listed. Click Start to proceed with the upgrade.
If an error occurs during the upgrade, an error message appears.
9 After the upgrade, check if the settings in the ChemStation Administration Tool
still match your original system settings before the upgrade.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 74
8 Upgrade Secure Workstation
Upgrade a Secure Workstation
10 If you disabled it before, enable Secure File I/O again after the upgrade is
finished.
a Click Start> All Programs> Agilent Technologies> ChemStation
Administration Tool to open the ChemStation Administration Tool.
b In the ChemStation Administration Tool, select the enable secure file IO
check box.
Existing instrument configuration can remain unchanged after the upgrade.
NOTE
The Classic drivers are not supported with ChemStation C.01.08 and later.
Please check “Convert Classic Instrument Drivers to RC.NET” on page 66 for
details.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 75
9 Uninstall the Software
About Uninstallation 76
Run the OpenLab CDS Uninstallation Wizard 77
This chapter contains information on the uninstallation by using the OpenLab
Uninstallation Wizard.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 76
9 Uninstall the Software
About Uninstallation
About Uninstallation
Like the installation, the uninstallation is automated by the OpenLab CDS
ChemStation Edition Installer.
For your convenience, the installer uses the same user interfaces for the
software uninstallation of all configurations (standalone or networked
workstation). The OpenLab Uninstallation Wizard is found under the Maintenance
section of the installer. It guides you through the uninstallation steps.
NOTE
Do not use the Windows uninstallation tool for uninstalling OpenLab CDS
ChemStation Edition!
NOTE
If the installer was not used for installation, any manually installed additional
software such as Headspace, PAL, or third party drivers must be uninstalled
using the Windows Control Panel before the core software can be uninstalled.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 77
9 Uninstall the Software
Run the OpenLab CDS Uninstallation Wizard
Run the OpenLab CDS Uninstallation Wizard
1 Select Start> All Programs> Agilent Technologies> Uninstall OpenLab CDS.
The OpenLab Uninstallation Wizard opens.
2 In the Shared Components screen, select the Uninstall Software Verification
and Uninstall PostgreSQL check box.
Note: Software Verification Tool needs to be uninstalled if you wish to
re-install OpenLab CDS ChemStation Edition at a later time.
3 In the Summary screen under Uninstallation of OpenLab CDS ChemStation
Components, there is a list of the components you want to uninstall.
4 Select Start to start the uninstallation.
If you want to abort the uninstallation, select Cancel. If you want to change
any settings, select Back.
All listed components are automatically uninstalled, one after another.
5 When the uninstallation has finished, click Finish to close the uninstallation
wizard.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 78
10 Troubleshooting
Reconfigure Instruments Using Classic Drivers After Upgrade to ChemStation
C.01.08 or Higher 79
The chapter gives some troubleshooting hints.
Secure Workstation for OpenLab CDS ChemStation Edition Installation 79
10 Troubleshooting
Reconfigure Instruments Using Classic Drivers After Upgrade to ChemStation
C.01.08 or Higher
Reconfigure Instruments Using Classic Drivers
After Upgrade to ChemStation C.01.08 or Higher
If you missed the recommendations from the driver preparation described in the
OpenLab CDS ChemStation Edition Upgrade Guide (CDS_CS-Upgrade.pdf),
ChemStation will start up, but instruments with the Classic driver will no longer
be available. You will receive a notice that you need to reconfigure your
instrument.
The classic driver instruments and methods are not updated to RC.NET
automatically. To adjust them, perform the following steps.
To check if a classic method is used, go to Windows Explorer and view the
contents of the corresponding method folder. The classic method will not have
the Agilent folder listed.
1 Record the IP address for each classic driver instrument.
2 Verify there is a backup of the methods and data to another location.
3 To reconfigure the instrument, select the instrument in the OpenLab Control
Panel.
4 In the ribbon, click Configure Instrument.
A message is displayed, requesting you to reconfigure your instrument.
Only with RC.NET driver methods
Secure Workstation for OpenLab CDS ChemStation Edition Installation 80
10 Troubleshooting
Reconfigure Instruments Using Classic Drivers After Upgrade to ChemStation
C.01.08 or Higher
5 Click OK.
The Configure Instrument panel opens. The Use classic drivers check box is no
longer shown.
6 To update the instrument to use the RC.net driver, select the instrument in the
Configurable Modules panel, then click the arrow to add it to the Selected
Modules panel.
7 Enter the IP Address. Only for 35900E A/D instrument: Click Get Serial Number
and Firmware to get the corresponding entries. For the example below the
serial number and firmware version are updated upon successful connection
to the 35900E A/D instrument.
This completes the setup of the instrument.
8 Click OK to load the instrument configuration.
9 Specify the Method Load on Startup option and select one of the following
options:
a Always ask the user to choose an option
b Download method to instrument on start up (select this option to match the
behavior of the classic driver)
c Upload method from instrument
d New method from instrument
Secure Workstation for OpenLab CDS ChemStation Edition Installation 81
10 Troubleshooting
Reconfigure Instruments Using Classic Drivers After Upgrade to ChemStation
C.01.08 or Higher
10 Click OK to complete.
11 Launch the newly configured instrument.
12 To convert a method to RC.NET driver, load the method. If a dialog like the
following opens, click OK.
Review the method, and if no updates are needed, add a comment such as
"Updated to RC.NET" to the method when saving it.
Methods are converted to RC.NET when loaded. After saving to disk,
converted methods have an additional Agilent folder and
RapidControl.InstrumentConfig file.
Only with RC.NET driver methods
Secure Workstation for OpenLab CDS ChemStation Edition Installation 82
11 Appendix
Change the PC Name 83
Sales and Support Assistance 84
Secure Workstation for OpenLab CDS ChemStation Edition Installation 83
11 Appendix
Change the PC Name
Change the PC Name
A change of the PC name may be scheduled to avoid duplicate names in a
network or can be due to a policy change. Internally, the software components
use localhost for the Workstation, so no additional action is required to continue
running the software. However, if you need to change a PC name after
installation, the license file for the PC will be no longer valid.
Instruments are licensed by installing files that are specifically created for a given
configuration. The text of the license files contains references to the computer
name and the hardware address of the network card. If one of the parameters is
changed, the license no longer applies.
You will need to generate and install a license file with the new PC name (host
name).
Secure Workstation for OpenLab CDS ChemStation Edition Installation 84
11 Appendix
Sales and Support Assistance
Sales and Support Assistance
Please check the following web site for your local sales and support contact:
https://www.agilent.com/en/support
www.agilent.com

Agilent Technologies Inc. 2014-2020
Published in Germany
05/2020
M8313-90004*
Part No: M8313-90004 Rev. D
In This Book
This installation guide provides instructions
to install and configure Agilent Secure
Workstation for OpenLab CDS ChemStation
Edition.