JANUARY 2022
APHL GUIDE FOR
CLIA INTERNAL AUDITS
RELATED TO HIGH COMPLEXITY TESTING
®
APHL CLIA Audit Checklist | 2
CONTENTS
Introduction ............................................ 3
Acknowledgments .................................3
APHL Workgroup .............................................................. 3
APHL Staff ....................................................................... 3
CDC ............................................................................... 3
CMS CLIA ......................................................................... 3
List of Assessors by CLIA Regulatory
Requirement ..........................................4
1. Facilities & Safety ............................5
2. Personnel .........................................7
2.1 General CLIA Personnel Requirements (Laboratory
Director Responsibilities Not Covered Elsewhere) .. 7
2.2 Qualifications ........................................................... 8
2.3 Responsibilities ....................................................... 9
2.4 Training .................................................................. 11
2.5 Competency Assessment ...................................... 12
3. Pre-analytic & Post-analytic
Systems ..........................................15
3.1 Pre-analytic Systems............................................. 15
3.2 Post-analytic .......................................................... 17
4. Analytic—Test Systems, Reagents,
Materials & Supplies ......................21
5. Analytic—Equipment Maintenance,
Function Checks & Calibration .....23
6. Analytic—Testing ...........................26
6.1 Procedure Manual ................................................. 26
6.2 Analytic Systems—Verification and Validation
(establishment of performance specifications) ..... 29
7. Analytic—Quality Control ..............31
7.1 QC—General .......................................................... 31
7.2 QC—Bacteriology .................................................. 36
7.3 QC—Mycobacteriology.......................................... 37
7.4 QC—Mycology ....................................................... 37
7.5 QC—Parasitology .................................................. 38
7.6 QC—Virology ......................................................... 38
7.7 QC—Other than Microbiology ............................... 38
8. Proficiency Testing (PT) &
Alternative Assessment ................39
8.1 PT—Enrollment ..................................................... 39
8.2 PT—Testing of Samples ........................................ 39
8.3 PT—Evaluation of Performance ............................ 41
8.4 Comparison of Test Results ................................... 42
9. General Laboratory Systems &
Retention Requirements ...............43
9.1 General Laboratory Systems §493.1230 ............... 43
10. Quality Assessment & Corrective
Action ..............................................44
10.1 Quality Assessment ............................................... 44
10.2 Corrective Action ................................................... 47
APHL CLIA Audit Checklist | 3
INTRODUCTION
This document was developed as a quality improvement tool for internal “self” assessments of CLIA laboratory activities.
Throughout the guide, State Operations Manual (SOM)/Code of Federal Regulations (CFR) numbers are hyperlinked within the
tables for a full explanation of each requirement. Access the full documents:
SOM: State Operations Manual Appendix C - Survey Procedures and Interpretive Guidelines for Laboratories and Laboratory
Services (Rev. 166, 02/03/17)
CFR: Electronic Code of Federal Regulations (Title 42: Public Health, Part 493 – Laboratory Requirements)
The information and content in this document reects the CFR requirements in place on March 31, 2021. APHL makes no
representations of any kind, express or implied about the completeness, accuracy, reliability, or suitability of this checklist with
respect to ongoing updates to reect amendments to the eCFR website. Therefore, the use of this document is strictly at your
own risk. However, the eCFR website link provided in this document is hyperlinked to the current eCFR website version.
ACKNOWLEDGMENTS
APHL would like to acknowledge the following individuals and organizations for their contributions, feedback and expertise to this
project:
APHL Workgroup
Hyder Al Janabi
Mary Bonifas, BS, CQA(ASQ)
Bonita Bryant, MT(ASCP)
Leann Covington
Gillian Edwards
Laurie Gregg, MT(ASCP)
Analiza Gross, MBA, MT(ASCP)
Lydia Mikhail
Laura Naught, PhD, NRCC(TC)
Rajesh Parmar
Kathy Ross
Deborah Severson, MT(ASCP)
Laboratory Systems and Standards committee chair
Kimyatta Smith, MS
Tamara Theisen, MT(ASCP)
APHL Staff
Kuki Hansen, BVMS, MPH, DACVPM
Lorelei Kurimski, MS
Tina Su, MPH
Andrea Wright
CDC
Nancy Anderson, MMSc, MT(ASCP)
Víctor R. De Jesús, PhD
Marie Earley, PhD
Meredith Korth, CQA (ASQ), Booz Allen Hamilton
Atis Muehlenbachs, MD, PhD, FCAP
Marranda S. Scott, MLS(ASCP)
Angela Thompson, MS, MT(ASCP)
CMS CLIA
APHL CLIA Audit Checklist | 4
LIST OF ASSESSORS BY CLIA REGULATORY REQUIREMENT
Category of CLIA Regulatory Requirements Assessed Assessor
1. Facilities & Safety
2. Personnel
2.1 General CLIA Personnel Requirements
2.2 Qualications
2.3 Responsibilities
2.4 Training
2.5 Competency Assessment
3. Preanalytic & Postanalytic Systems
3.1 Preanalytic Systems
3.2 Postanalytic Systems
4. Analytic—Test Systems, Reagents, Materials & Supplies
5. Analytic—Equipment Maintenance, Function Checks & Calibration
6. Analytic—Testing
6.1 Procedure Manual
6.2 Verication and Validation
7. Analytic—Quality Control
8. Prociency Testing & Alternative Assessment
8.1 Enrollment
8.2 Testing of samples
8.3 Evaluation of performance
8.4 Comparison of test results
9. General Laboratory Systems & Retention Requirements
10. Quality Assessment & Corrective Action
10.1 Quality Assessment
10.2 Corrective Action
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Facilities & Safety | Page 5
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
1. FACILITIES & SAFETY
# SOM/CFR* Requirement Examples of Compliance Notes
1.1
D3001
§493.1101(a)(1)
Does the laboratory have adequate
space, ventilation and utilities necessary
for conducting all phases of the testing
process?
Make laboratory floor plans available.
Observe during laboratory tours
(space: crowded; ventilation: check air
flow strips; adequate lighting; fumes;
availability of electrical outlets).
Ask about backup power.
Arrange work areas to minimize
problems in specimen handling,
examination and testing, and the
reporting of test results.
Place instruments, equipment and
computer systems in locations where
their operation is not adversely
affected by physical or chemical
factors, such as heat, direct sunlight,
vibrations, power fluctuations.
Requirement met?
Yes
No
Notes:
1.2
D6083, D6084
§493.1445(e)(2)
Laboratory Director Responsibility
(cannot be delegated):
Does the Laboratory Director ensure that
the physical plant and environmental
conditions of the laboratory:
Are appropriate for the testing
performed?
Provide a safe environment in which
employees are protected from physical,
chemical, and biological hazards?
Assure there are quality systems in
place.
Ask about any safety incident(s) in the
laboratory and what was done.
Requirement met?
Yes
No
See CLIA Personnel Review Worksheet Tool.
Notes:
1.3
D3003
§493.1101(a)(2)
Is the laboratory space maintained to
minimize contamination of specimens,
equipment, instruments, reagents,
materials and supplies?
Observe during the lab tour:
Lab is clean and uncluttered
Eyewash or hand wash facilities are
accessible
Ventilation is clear
Power strips are acceptable to use, but
extension cords are not
Lab bench decontamination log is
available.
Requirement met?
Yes
No
Notes:
1.4
D3005
§493.1101(a)(3)
Is lab space used for molecular
amplification procedures arranged to
provide unidirectional workflow with
separate areas for:
Specimen prep, amplification and
detection?
Reagent preparation (as applicable)?
Mechanism to detect cross-
contamination of patient specimens?
Observe unidirectional workflow (no
workarounds)
Environmental swipe test and blank for
contamination
Evidence of blank controls (NTCs) from
QC logs or run sheets.
Requirement met?
Yes
No
Notes:
1.5
D3007
§493.1101(b)
Does the laboratory have appropriate
and sufficient equipment, instruments,
reagents, materials and supplies for the
type and volume of testing performed?
Observe during lab tour or assess
through staff discussions about essential
laboratory equipment, (e.g. autoclave,
centrifuge, scales, correct reagents
for instrumentation, equipment list,
inventory list, etc.)
Is there a purchasing process in place
and is it sufficient?
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Facilities & Safety | Page 6
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
1.6
D3009
§493.1101(c)
The laboratory must be in compliance
with applicable federal, state and local
laboratory requirements.
Ask for the current certificates or state
licensure documents.
Ask for the laboratory’s policy on
compliance with these requirement
even if there is no state licensure or
certificates.
Requirement met?
Yes
No
Notes:
1.7
D3011
§493.1101(d)
Are safety procedures established,
accessible and observed to ensure
protection from physical, chemical,
biochemical and electrical hazards, and
biohazardous materials?
Check that findings from any laboratory’s
safety survey report have been corrected.
Laboratory specific documentation—
including the biosafety manual, waste
management plan, chemical hygiene
plan and, if applicable, the laboratory’s
bloodborne pathogen exposure control
plan and training documents—are
reviewed during the annual safety survey.
Requirement met?
Yes
No
This includes all types of laboratories
(e.g., permanent, mobile, temporary, etc.)
Notes:
1.8
D3027, 3029,
3031, 3033,
3037, 3039, and
3041
§493.1101(e)
Are records and, as applicable, slides,
blocks, and tissues maintained and
stored under conditions that ensure
proper preservation?
How are records stored? What is your
record retention policy? What is your
policy for the temperature and humidity
control for the records? What is your off-
site archive and retrieval policy?
Check state requirements as applicable—
go with most stringent.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Personnel | Page 7
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
2. PERSONNEL
2.1 General CLIA Personnel Requirements (Laboratory Director Responsibilities
Not Covered Elsewhere)
Refer to the CLIA Personnel Review Worksheet Tool for more information on each of the following requirements/responsibilities:
# SOM/CFR* Requirement Examples of Compliance Notes
2.1.a
D6079
§493.1445
(a)(b)
Note:
Also covered in
2.3.c
Is the laboratory director responsible for
the overall operation and administration
of the laboratory, including:
Employment of personnel who are
competent to perform test procedures,
record and report test results
promptly, accurately and proficiently,
and for assuring compliance with the
applicable regulations?
Delegation of qualified personnel
to perform the duties of Technical
Supervisor, Clinical Consultant,
General Supervisor, and Testing
Personnel?
Ensuring all duties delegated are
properly performed?
If a new CLIA director:
Ask for educational records,
competency records, transcripts,
certificates, diplomas, laboratory
organizational charts, etc.
Identify technical supervisors and
policy on delegation of duties.
Requirement met?
Yes
No
Notes:
2.1.b
D6080
§493.1445(c)
Note:
D6144, D6145
and D6146
§493.1463(a)
(1-2) are covered
in 2.3.e
Is the laboratory director accessible
to the laboratory to provide onsite,
telephone or electronic consultation as
needed?
Documentation addressed in other
sections.
Requirement met?
Yes
No
Notes:
2.1.c
D6082
§493.1445
(e)(1)
Cannot be delegated:
Does the laboratory director ensure
that testing systems developed and
used for each of the tests performed
in the laboratory provide quality
laboratory services for all aspects of
test performance, which includes the
pre-analytic, analytic, and post-analytic
phases of testing?
Documentation addressed in other
sections.
Requirement met?
Yes
No
Notes:
2.1.d
D6100
§493.1445
(e)(10)
Cannot be delegated:
Does the laboratory director ensure
that a general supervisor provides
on-site supervision of high complexity
test performance by testing personnel
qualified under §493.1489(b) (4)?
Documentation addressed in other
sections.
Requirement met?
Yes
No
Notes:
2.1.e
D6101
§493.1445
(e)(11)
Cannot be delegated:
Does the laboratory director ensure the
employment/assignment of a sufficient
number of laboratory personnel with
appropriate education and either
experience or training to provide
appropriate consultation, properly
supervise and accurately perform tests
and report test results?
Documentation addressed in other
sections.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Personnel | Page 8
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
2.2 Qualifications
# SOM/CFR* Requirement Examples of Compliance Notes
For each person acting in the following
personnel roles, are qualifications
documented?
Since the previous internal audit/
inspection, the items below must be
current for new employees:
Diploma indicating degree in a
chemical, physical, biological,
clinical laboratory science or medical
technology, if applicable
Transcript, if diploma does not
specify area of study (transcripts
were requested by CMS inspectors
for degrees in public health, zoology,
epidemiology, forestry and veterinary
medicine)
Foreign degree equivalency if
educational requirement is from a
foreign institution
Professional training records
Consult regulations for specifics.
Requirement met?
Yes
No
Notes:
2.2.a
D6078
§493.1443
Laboratory Director Certificate or (state) license to indicate
qualification to direct the laboratory,
degree (MD or PhD) or other additional
certifications.
Requirement met?
Yes
No
See CLIA Personnel Review Worksheet Tool
Notes:
2.2.b
D6111
§493.1449
Technical Supervisor The laboratory director can also be the
technical supervisor.
Someone with a BS and MS degree with
years of experience can be the technical
supervisor with written designation.
Requirement met?
Yes
No
See CLIA personnel forms:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
2.2.c
D6135
§493.1455
Clinical Consultant MDs: License to practice medicine
from any US state or territory or board
certification by a CMS-approved board.
Note: No further documentation is
required (no need to produce foreign
educational equivalency), but the medical
license must be current and active (must
not be expired).
PhDs: Degree in a chemical, physical,
biological or clinical laboratory science,
plus active certification by a HHS-
approved board (e.g., ABMM or ABB).
Note: Further documentation may be
needed to produce foreign educational
equivalency of the PhD.
Requirement met?
Yes
No
See CLIA personnel forms:
CLIA Personnel Review Worksheet Tool
Clinical Consultant Personnel Form
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Personnel | Page 9
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
2.2.d
D6143
§493.1461
General Supervisor Documentation of two years laboratory
training or experience is required.
Requirement met?
Yes
No
See CLIA personnel forms:
CLIA Personnel Review Worksheet Tool
General Supervisor Personnel Form
Notes:
2.2.e
D6171
§493.1489
Testing Personnel Documentation of a degree (doctoral,
master’s or bachelor’s) in a chemical,
physical, biological or clinical laboratory
science, or medical technology from an
accredited institution.
Requirement met?
Yes
No
Notes:
2.3 Responsibilities
# SOM/CFR* Requirement Examples of Compliance Notes
2.3.a
D6107
§493.1445
(e)(15)
Has the laboratory director specified in
writing the responsibilities and duties
for all:
Clinical consultants?
Technical supervisors?
General supervisors?
Testing personnel?
Written documentation, policies,
procedures or forms for all CLIA
personnel signed by the laboratory
director or designee:
Clinical consultants
Technical supervisors
General supervisors
Testing personnel
Types of documentation:
Position descriptions
Competency documents
Training records
SOPs and authorizations
Requirement met?
Yes
No
See CLIA personnel forms:
Clinical Consultants
Technical Supervisors
General Supervisors
Testing Personnel
Supervisor Competency Assessment
Procedure
CLIA Supervisory Competency
Assessment for General Supervisor Form
CLIA Supervisory Competency
Assessment for Clinical Consultant Form
CLIA Supervisory Competency
Assessment for Technical Supervisor
Form
Notes:
2.3.b
D6107
§493.1445
(e)(15)
For testing personnel, is there
documentation that specifies:
Testing procedures the individual is
authorized to perform?
Whether supervision is required
for specimen processing, test
performance or result reporting?
Whether supervisory or director review
is required prior to reporting patient
test results?
Written documentation, policies,
procedures or forms for all testing
personnel signed by the laboratory
director or designee indicating the tests
the individual is authorized to perform
without supervision
Note: supervisory review is required
before patient test results are reported
for all testing personnel.
Types of documentation:
Position descriptions
Competency documents
Training records
SOPs and authorizations
Requirement met?
Yes
No
See CLIA Testing Personnel Form
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Personnel | Page 10
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
2.3.c
D6079
§493.1445
(a)(b)
Note: Also
covered in 2.1.a
Laboratory Director Responsibility:
Is the laboratory director responsible for
the overall operation and administration
of the laboratory, including:
Employing personnel who are
competent to perform test procedures;
record and report test results promptly,
accurately and proficiently; and
assure compliance with the applicable
regulations?
Delegating qualified personnel to
perform the duties of technical
supervisor, clinical consultant, general
supervisor and testing personnel?
Ensuring all duties delegated are
properly performed?
Written documentation of
reapportionments.
Requirement met?
Yes
No
See CLIA Personnel Review Worksheet Tool
Notes:
2.3.d
D6136, D6137,
D6138 and
D6140
§493.1457
(a)(b)(d)
Clinical Consultant Responsibility:
Does the laboratory have a clinical
consultant available to provide:
Consultation to submitters?
Assist submitters to ensure
appropriate test are ordered?
Communicate with submitters about
the quality of test results reported and
their interpretation concerning specific
patient conditions?
Communication log that shows the
clinical consultant provides consultation
with submitters.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Clinical Consultant Personnel Form
Notes:
2.3.e
D6144, D6145
and D6146
§493.1463
(a)(1-2)
Note: Also
covered in 2.1.b
General Supervisor Responsibility:
Does the general supervisor:
Provide day-to-day supervision of high
complexity test performance?
Remain accessible to testing
personnel at all times testing
is performed to provide on-site,
telephone or electronic consultation
to resolve technical problems
in accordance with policies and
procedures established either by
the laboratory director or technical
supervisor?
Evidence of day-to-day review of records,
such as QC record or organizational
charts that display reporting lines.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
General Supervisor Personnel Form
Notes:
2.3.f
D6080
§493.1445(c)
Note: Also
covered in 2.1.a
Laboratory Director Responsibility:
Is the laboratory director accessible
to the laboratory to provide onsite,
telephone or electronic consultation as
needed?
Overall review of system compliance.
Requirement met?
Yes
No
See CLIA Personnel Review Worksheet Tool
Notes:
2.3.g
D6174
§493.1495(a)
Testing Personnel Responsibility:
Is there documented evidence that
testing person-nel limit patient testing to
only those tests which they have been
authorized to per-form by the laboratory
director?
Review test results to ensure performed
by an authorized testing personnel, or
train-ing and competency assessment
records.
Requirement met?
Yes
No
See CLIA Personnel Review Worksheet Tool
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Personnel | Page 11
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
2.4 Training
# SOM/CFR* Requirement Examples of Compliance Notes
2.4.a
D6102
§493.1445
(e)(12)
Technical Supervisor Responsibility
(Delegated by LD):
Does the technical supervisor ensure
that, prior to testing patients’ specimens,
all personnel:
Have the appropriate education and
experience?
Receive the appropriate training for
the type and complexity of services
offered?
Have demonstrated that they can
perform all testing operations reliably
to provide and report accurate results?
Personnel qualification records
(e.g., transcripts, diplomas, foreign
evalutions, equivalency records,
training records, competency
assessment records, etc.).
LD or designee indicates TS
responsibilities in writing.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
2.4.b
D6120
§493.1451
(b)(7)
Technical Supervisor Responsibility:
Does the technical supervisor ensure
that staff training needs are identified
and that all testing personnel receive
regular in-service training and education?
Note: This responsiblility can be delegated
in writing to GS (see 2.4.c).
Personnel training records.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
2.4.c
D6151
§493.1463
(b)(3)
General Supervisor Responsibility
(If delegated by LD or TS):
Does the general supervisor provide
orientation to all testing personnel?
Personnel training records, onboarding
checklist.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
General Supervisor Personnel Form
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Personnel | Page 12
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
2.5 Competency Assessment
# SOM/CFR* Requirement Examples of Compliance Notes
2.5.a
D6087
§493.1445
(e)(3)(iii)
Technical Supervisor Responsibility
(Delegated by LD): Does the technical
supervisor ensure that laboratory
personnel are performing the test
methods as required for accurate and
reliable results?
Note: This responsibility can be delegated
in writing to GS (see 2.5.e).
Competency assessment (CA) records
Training records
Quality assurance records
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
CLIA Testing Personnel Test System
Authorization Form
Notes:
2.5.b
D6103
§493.1445
(e)(13)
Technical Supervisor Responsibility
(Delegated by LD): Does the technical
supervisor:
Ensure that policies and procedures
are established for monitoring
individuals who conduct pre-analytical,
analytical and post-analytical phases
of testing to assure that they are
competent and maintain their
competency to process specimens,
perform test procedures and report
test results promptly and proficiently?
Identify needs for remedial training or
continuing education to improve skills?
Note: This responsibility can be delegated
in writing to GS (see 2.5.e).
CA records
Policy and procedure documents
Documents where staff have
acknowledged that they understand
the policies and procedures (signature
page—digitally, electronically, manually,
attestation, etc.)
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
CLIA Testing Personnel Test System
Authorization Form
Notes:
2.5.c
D6120
§493.1451
(b)(8)
Technical Supervisor Responsibility:
Does the technical supervisor ensure
policies and procedures are established
to evaluate the competency of all
testing personnel and assure that staff
maintain their competency to perform
test procedures and report test results
promptly, accurately and proficiently?
Note: This responsibility can be delegated
in writing to GS (see 2.5.e).
CA records:
Are easy to follow, organized and
standardized
Have a summary of the schedule for
all personnel in the front of the CA
records (recommended).
Review policies and procedures
related to compentencies.
Competency records.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
CLIA Testing Personnel Test System
Authorization Form
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Personnel | Page 13
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
2.5.d
D6121–D6126
§493.1451
(b)(8)(i-vi)
Technical Supervisor Responsibility:
Do competency assessments include
each of the following methods:
Direct observation of testing?
Monitoring the recording and reporting
of results?
Review of intermediate results, QC, PT
and PM?
Direct observation of maintenance and
function checks?
Assessment of test performance
through repeat, blind or external PT
testing?
Assessment of problem solving skills?
Note: This responsibility can be delegated
in writing to GS (see 2.5.e).
CA records:
Are performed by an individual who
has supporting documents and
qualifications to do so (either GS or TS)
Have details of what test was
assessed, including dates for each of
the six methods of assessment
Include the results of test performance
and evaluation including worksheets
(if PT was used, specifics for PT event
must be included or a copy of the PT
event records)
Show evidence of supervisory review
Include documentation of successful
retraining as necessary
Note: All six competencies must be
assessed within the same calendar year,
but not necessarily at the same time.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
2.5.e
D6151
§493.1463
(b)(4)
General Supervisor Responsibility
(if delegated by LD or TS):
Does the general supervisor annually
evaluate and document the performance
of all testing personnel.
Annual competency assessment records,
to include dates.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
General Supervisor Personnel Form
CLIA Testing Personnel Test System
Authorization Form
Notes:
2.5.f
D6179
§493.1495
(b)(5)
Testing Personnel Responsibility:
Is there documented evidence that
testing per-sonnel are capable of
identifying problems that may adversely
affect test system per-formance?
Corrective action reports
Event logs
Root cause analysis
Instrument logs
Note: Documentation may be used as
evidence for competency assessment.
Requirement met?
Yes
No
See CLIA Personnel Review Worksheet Tool
Notes:
2.5.g
D6175
§493.1495
(b)(1)
Testing Personnel Responsibility:
Is there documented evidence that
testing personnel follow laboratory
procedures for:
Specimen handling and processing?
Test analyses?
Reporting and maintaining records of
patient test results?
Review of run reports with raw data,
sample rejection logs, quality control
logs, corrective action reports, review of
result reports.
Note: Documentation may be used as
evidence for competency assessment.
Requirement met?
Yes
No
See CLIA Personnel Review Worksheet Tool
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Personnel | Page 14
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
2.5.h
D6177
§493.1495
(b)(3)
Testing Personnel Responsibility:
Is there documented evidence that
testing personnel adhere to quality
control policies and document all
quality control activities, instrument and
procedural calibrations and maintenance
performed?
Quality control records
Instrument records
Maintenance records
Documents where staff have
acknowledged that they understand
the policies and procedures (signature
page—digitally, electronically, manually,
attestation, etc.)
Note: Documentation may be used as
evidence for competency assessment.
Requirement met?
Yes
No
See CLIA Personnel Review Worksheet Tool
Notes:
2.5.i
D6127–D6129
§493.1451
(b)(9)
Technical Supervisor Responsibility:
Are testing personnel competency
assessments conducted and
documented:
Semiannually during the first year the
individual tests patient specimens?
Annually thereafter?
When a new method or instrument
change is introduced, individual
is reevaluated to include the use
of the new test methodology or
instrumentation prior to reporting
patient test results?
Competency assessment records—may
be delegated in writing to GS (see 2.5.e).
Check for new testing personnel CLIA
testing start date and ensure the
laboratory has records of two compe-
tency assessments during the 12
months following this start date, with
at least one assessment within the
first six months.
Check existing personnel records for
new methods authorized to perform.
If the testing methodology changed,
retraining will be needed at least once
and documented as necessary.
Ensure there are records of annual
competency assessment for existing
personnel.
Records must show evidence of
supervisory review.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
2.5.j
D5209
§493.1235
As specified in the personnel
requirements in Subpart M—Personnel
for Nonwaived Testing, the laboratory
must establish and follow written policies
and procedures to assess employee and,
if applicable, consultant competency.
Policies or procedures in place to
assess personnel performing pre-
analytical (accessioning), analytical or
post-analytical (reporting)
Documents where staff have
acknowledged that they understand
the policies and procedures (signature
page—digitally, electronically, manually,
attestation, etc.)
Requirement met?
Yes
No
See CLIA Testing Personnel Test System
Authorization Form
Notes:
Additional supporting documents for this section:
Supervisor Competency Assessment Procedure
CLIA Supervisory Competency Assessment for General Supervisor Form
CLIA Supervisory Competency Assessment for Clinical Consultant Form
CLIA Supervisory Competency Assessment for Technical Supervisor Form
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Pre-analytic & Post-analytic Systems | Page 15
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
3. PRE-ANALYTIC & POST-ANALYTIC SYSTEMS
3.1 Pre-analytic Systems
# SOM/CFR* Requirement Examples of Compliance Notes
3.1.a
D5301
§493.1241(a)
Does the laboratory require written or
electronic requests for patient testing?
Specimen submission/requisition
form should be available for specimen
testing.
Quality assurance manual.
Policy or standard operating procedure
addressing the need for orders (how
long does the lab have to order?, turn
around time, etc.).
Requirement met?
Yes
No
Notes:
3.1.b
D5303
§493.1241(b)
If the laboratory accepts oral requests,
are written or electronic confirmation
requests received within 30 days?
This may be stated in a policy or
procedure for specimen receipt and
handling, or the lab may have records
of communication that provide
evidence of compliance.
Quality assurance manual.
Policy or standard operating procedure
addressing the need for orders (how
long does the lab have to order?, turn
around time, etc.).
Requirement met?
Yes
No
Notes:
3.1.c
D5305
§493.1241
(c)(1-6, 8)
Do test requisitions contain the following
information?
Information necessary to identify
the authorized submitter to send
test results and if appropriate, an
individual contact person to enable
reporting of imminently life threatening
results or panic or alert values?
Patient’s name or unique identifier?
The sex and age or date of birth of the
patient?
The test(s) to be performed?
The specimen source, when
appropriate?
Date and, if appropriate, time of
specimen collection?
Any additional information relevant/
necessary for a specific test?
Review a sampling of specimen
submission/requisition forms for
required information.
The laboratory may use a
supplemental form to collect
information relevant for specific tests.
State Operations Manual: Does the
laboratory have a policy/procedure
for what to do when a requisition
is missing required information?
Laboratories must either obtain the
missing information, or report results
and indicate on the test report any
limitations of test results due to the
omission of patient information (See
page 1 of Appendix C for an inspector
and laboratory guide).
Requirement met?
Yes
No
Notes:
3.1.d
D5309
§493.1241(e)
Does the laboratory ensure that test
requisitions transcribed into a record
system (paper or electronic) are entered
accurately?
QA monitors in place for requisition
accessioning; the laboratory is
responsible to ensure data is entered
correctly and must have a procedure
to check data entered. This may be
in a policy or procedure for specimen
receipt and handling.
Demographic change logs, corrective
report logs, QA monthly assessment
or quarterly report, assessment using
LIMS print out, pull metrics from
double data entry from LIMS (error
rates).
Requirement met?
Yes
No
See Demographic Change Log Example
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Pre-analytic & Post-analytic Systems | Page 16
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
3.1.e
D5311
§493.1242
(a)(1-8)
Does the laboratory establish and follow
written policies and procedures for each
of the following, if applicable:
Patient preparation?
Specimen collection?
Specimen labeling, including patient
name or unique identifier and, when
appropriate, specimen source?
Specimen storage and preservation?
Conditions for specimen
transportation?
Specimen processing?
Specimen acceptability and rejection?
Specimen referral?
Information in a test directory should
be consistent with laboratory test
procedures for the following:
Acceptable sample/specimen type for
testing
Storage and preservation of specimen
prior to shipping
Transport medium
Specimen labeling
Shipping instructions which include
specimen handling requirements
Documentation:
Test requisitions
Test menus
SOPs (ensure SOPs include
acceptability and rejection)
Orders
Containers
At least two unique patient identifiers
on specimen container labeling,
proper identification of patient
Requirement met?
Yes
No
Note: Crosswalk all policies and procedures
to all physical evidence.
Notes:
3.1.f
D5313
§493.1242(b)
Does the laboratory document the date
and time of specimen receipt?
Date and time packages/samples are
received is documented (e.g, requisition
forms, sample receipt logs, LIMS sample
logs, etc.).
Requirement met?
Yes
No
Notes:
3.1.g
D5315
§493.1242(c)
Does the laboratory refer specimens only
to other CLIA-certified laboratories?
Documentation:
Reference lab CLIA certificates
Specified in SOP or procedure
Requirement met?
Yes
No
Notes:
3.1.h
D5317
§493.1242
(a)(1-8)
Does the laboratory provide clients
written instructions to meet the
laboratory’s requirements for the
information outlined in 3.1.e?
Test menu/test directory
SOPs, policies
Requirement met?
Yes
No
Notes:
3.1.i
D3027
§493.1105
(a)(1)
Record Retention: Are test requisitions
and authorizations retained for at least
two years?
Note: Longer retention times might be
in place for other state, local, federal
regulations and must also be followed. The
CLIA two-year rule must be followed at a
minimum.
Test requisistions available for at least
two years (can be longer)—address
discontinued procedures as well
SOPs, policies
Check state requirements as
applicable—go with most stringent
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Pre-analytic & Post-analytic Systems | Page 17
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
3.1.j
D6138
§493.1457(b)
Clinical Consultant Responsibility:
Does the laboratory have a clinical
consultant available to assist submitters
in ensuring appropriate tests are ordered
to meet the clinical expectations?
A CC, other than the LD, is identified.
Note: CC can also be the LD.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Clinical Consultant Personnel Form
Notes:
3.1.k
D6138, D6140
§493.1457
(b)(d)
Clinical Consultant Responsibility:
Is the CC available to assist the
laboratory’s clients in ensuring that
appropriate tests are ordered to meet the
clinical expectations?
CC: Inquire into records of
communication between CC and
submitter on specific patient test
requests or test results (as applicable).
Note: CC can also be the LD.
Requirement met?
Yes
No
Notes:
3.2 Post-analytic
# SOM/CFR* Requirement Examples of Compliance Notes
3.2.a
D5801
§493.1291
(a)(1-3)
Does the laboratory have an adequate
manual or electronic system(s) in place
to ensure test results and other patient-
specific data are accurately and reliably
sent from the point of data entry to final
report destination in a timely manner?
Including the following:
Results reported from calculated data
Results and patient-specific data
electronically reported to network or
interfaced systems
Manually transcribed or electronically
transmitted results and patient-
specific information
Documentation of result reporting quality
checks, such as:
Supervisory review of reports (can be
LIS manager)
Audit log of electronic reports
Patient test managing report
Requirement met?
Yes
No
See Patient Test Management Evaluation
Form
Notes:
3.2.b
D5803
§493.1291(b)
Is test report information readily available
to the laboratory and to CMS or a CMS
agent upon request?
Request a random sampling of patient
testing records representing each test
system performed and corrected report
examples if available. Records should
include the specimen requisition, testing
worksheets/data printouts and the final
test report. The laboratory must be able
to retrieve test reports requested during
the CMS inspection.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Pre-analytic & Post-analytic Systems | Page 18
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
3.2.c
D58051
§493.1291
(c)(1-7)
Does the patient test report include the
following:
Patient’s name and identification
number or a unique identifier and
identification number?
Name and address of the laboratory
location where the test was
performed?
Test report date?
Test performed?
Specimen source, when appropriate?
Test result and, if applicable, the units
of measurement or interpretation, or
both?
Any information regarding the
condition and disposition of specimens
that do not meet the laboratory’s
criteria for acceptability?
Review the sampling of patient testing
records (3.2.b.) requested for required
information.
If any of the test results were from a
laboratory outside of your laboratory’s
CLIA certificate, check that this is
noted on the report.
Reports should include a disclaimer, if
applicable. For example, test reports
for laboratory developed tests must
include a disclaimer indicating “The
performance characteristics of this
test were determined by (Laboratory
Name). It has not been cleared or
approved by the US Food and Drug
Administration.”
Include additional patient identifiers,
ordering physician and date/time of
specimen collection, if available.
Requirement met?
Yes
No
Notes:
3.2.d
D6139
§493.1457(c)
Clinical Consultant Responsibility:
Has the clinical consultant reviewed
patient report templates to ensure
the test results include pertinent
information required for specific patient
interpretation?
Documentation of review test report
templates.
Documentation that CC has reviewed
reports for each type result reported to
include any interpretation, comments or
disclaimers and test report results are
clear and easy to understand.
Example documentation:
Mock up test report
List of disclaimer
Interpretation comments
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Clinical Consultant Personnel Form
Notes:
3.2.e
D6098
§493.1445
(e)(8)
Clinical Consultant Responsibility
(If delegated by LD): Does the clinical
consultant ensure that reports of test
results include pertinent information
required for interpretation?
See evidence for 3.2.d.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Clinical Consultant Personnel Form
Notes:
3.2.f
D5807
§493.1291(d)
Are pertinent reference intervals or
normal values, as determined by the
laboratory performing the test, available
to individuals ordering the test or
responsible for using the test results?
Note: The result of test performed by
another laboratory must reflect the
performing laboratory’s reference or normal
range, and other interpretation comments.
Reference interval (normal value)
information should be included in each
test procedure (see 6.1.b).
Examples:
List of reference intervals
Reports with reference intervals
SOPs should include reference
intervals—if they don’t, must have a list
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Pre-analytic & Post-analytic Systems | Page 19
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
3.2.g
D5809
§493.1291(e)
Does the laboratory have a means to
provide clients:
Test methods?
Performance specifications
established or verified by the
laboratory if applicable?
Information on interpretation of test
results including interferences?
Are pertinent updates on testing
information provided to clients whenever
changes occur that affect test results or
interpretation of test results?
A test directory
SOPs
Validation/verification data
Emails or letters to clients regarding
updates
Requirement met?
Yes
No
Notes:
3.2.h
D6099
§493.1445
(e)(9)
Clinical Consultant Responsibility
(If delegated by LD): Does the clinical
consultant ensure that consultation is
available to the laboratory’s clients on
matters relating to the quality of the test
results reported and their interpretation
concerning specific patient conditions?
If delegated to CC in writing by the LD:
Look through communication logs for
examples
Complaints or lack of complaints
Letter from clinical consultant stating
hours and contact information
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Clinical Consultant Personnel Form
Notes:
3.2.i
D5811
§493.1291(f)
Are test results released only to
authorized persons and, if applicable, the
individual responsible for using the test
results and the laboratory that initially
requested the test?
Policy statement
Review data entry if this is auto
reported
Requirement met?
Yes
No
Notes:
3.2.j
D5813
§493.1291(g)
Does the laboratory immediately alert
the individual or entity requesting the
test and, if applicable, the individual
responsible for using test results when
any test result indicates an imminently
life-threatening condition, or panic or
alert value?
Procedure for reporting alert or panic
values should include:
Definition for laboratory test results
that have been determined to indicate
an imminently life-threatening
condition, or panic or alert value
How alert values will be communicated
To whom alert values may be
communicated (role/qualifications)
Records should include:
Date, time and to whom test results
were reported
Method used to communicate (phone,
email, other)
Acknowledgement of receipt
The laboratory should have a
communication log of alert values.
Requirement met?
Yes
No
Notes:
3.2.k
D5815
§493.1291(h)
Has the laboratory established
timeframes for reporting test results?
If there is a delay in urgent or timely
requests, does the laboratory notify
appropriate individuals?
The test directory should include
turnaround times.
Records of submitter notification based
on how long delay is and the impact on
patient care (e.g, documentation in the
laboratory information system, call log of
the notification, etc.).
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Pre-analytic & Post-analytic Systems | Page 20
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
3.2.l
D5815
§493.1291
(i)(1-3)
If a laboratory refers patient specimens
for testing, does the laboratory ensure
the following requirements are met:
Results or information directly
related to the interpretation of results
provided by the testing laboratory are
not revised by the referring laboratory.
If the testing laboratory is permitted
to send test results directly to the
authorized person who initially
requested the test, the referring
laboratory must retain or be able to
produce an exact duplicate of each
testing laboratory’s report.
The authorized person who initially
requested the test must be notified by
the referring laboratory of the name
and address of the laboratory where
testing was performed.
Evidence that you verified reference
laboratory’s CLIA number (via
certificate)
Copy of all the original reports from the
reference laboratories
Final report must include testing
laboratory’s name, address and CLIA
number
Test report does not require CLIA
number
Requirement met?
Yes
No
Notes:
3.2.m
D5815
§493.1291(j)
Are all test reports or records of the
information on the test report maintained
by the laboratory in a manner that
permits ready identification and timely
accessibility?
Test reports and records maintained
are easily identified and accessible.
Examples:
Copies of the reports
Demonstrate retrieval of reports
Identification and accessibility of storage
areas (only appropriate staff have
access).
Requirement met?
Yes
No
Notes:
3.2.n
D5815
§493.1291
(k)(1-3)
When errors in the reported patient test
results are detected, are the following
accomplished:
Authorized person ordering the test
and, if applicable, the individual using
the test results are promptly notified?
Prompt issuance of corrected
report(s)?
Duplicates of the original and
corrected reports are maintained?
Review corrected report examples from
the requested sampling of patient testing
records (3.2.b).
State Operations Manual: “Corrected
reports, either hard copy or electronic,
must clearly indicate both the corrected
result(s) and the fact that the report is a
corrected report.”
In the event that a corrected report was
issued, the following documentation is
required:
Non-conforming event (NCE) logs or
documentation
Documentation of notification
Copies of original reports are
maintained
Requirement met?
Yes
No
Notes:
3.2.o
D3041
§493.1105
(a)(6)
Record Retention: Are test reports
(preliminary, final and corrected/
amended) retained for at least two years?
Note: Longer retention times might be in
place for other state, local and federal
regulations and must also be followed. The
CLIA two-year rule must be followed at a
minimum.
Paper or electronic copy of original report
that contains the exact information as
sent to submitter, such as:
Destruction records
Record retention policy or procedure
Check state requirements as applicable –
go with most stringent.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Test Systems, Reagents, Materials & Supplies | Page 21
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
4. ANALYTIC—TEST SYSTEMS, REAGENTS, MATERIALS & SUPPLIES
# SOM/CFR* Requirement Examples of Compliance Notes
4.1
D5411
§493.1252(a)
Are test systems selected by the
laboratory?
Are tests performed following the
manufacturer’s instructions and in a
manner that provides test results within
the laboratory’s stated performance
specifications?
Recent review of package insert
SOPs
SOM lists specific parameters for:
Syphilis serology (antigen volume,
incubation time and temperature,
light source, rotator speed and
circumference, and conjugate titer)
International Normalized Ratio
calculation for coagulation testing
Requirement met?
Yes
No
Notes:
4.2
D5413
§493.1252
(b)(1-4)
Has the laboratory defined criteria for
conditions that are essential for proper
storage of reagents and specimens,
accurate and reliable test system
operation, and test result reporting?
Are the criteria consistent with the
manufacturer’s instructions, as
applicable?
Are the conditions monitored and
documented?
If specific conditions apply, do they
include:
Water quality?
Temperature?
Humidity?
Protection of equipment/
instruments from electrical
fluctuations?
Observe during lab tour:
Records of reagent and media
storage and handling consistent with
manufacturer’s instructions
Monitoring records for refrigerator,
freezer and room temperature
(room temperature must be defined
according to method)
Water quality reports if using water
from a water purification system for
testing procedures
Humidity if a test procedure or
reagent storage has specific humidity
requirements
Evidence and maintenance of
uninterruptible power supplies and
power strips or surge protectors
Requirement met?
Yes
No
Notes:
4.3
D5785
§493.1282
(b)(3)
Does the laboratory document corrective
actions taken when the criteria for proper
storage of reagents and specimens are
not met?
Temperature monitoring records
indicate whether the temperature
observed is within the defined
acceptable range
If temperatures are out of range, there
is documentation of corrective action
(e.g., reagents or specimens moved to
another refrigerator)
Corrective action log
NCE logs
Policy/SOP for storage of reagents and
specimens
Requirement met?
Yes
No
Notes:
4.4
D5415
§493.1252
(c)(1-4)
Are reagents, solutions, culture media,
control materials, calibration materials
and other supplies, as appropriate,
labeled to indicate:
Identity?
Titer, strength or concentration, if
applicable?
Storage requirements?
Preparation and expiration dates?
Other pertinent information required
for proper use?
Observe during lab tour:
Name
Titer, strength or concentration for
stocks and working solutions
Preparation date or opened date
Expiration date
Storage requirements
SOP
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Test Systems, Reagents, Materials & Supplies | Page 22
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
4.5
D5417
§493.1252(d)
Does the laboratory prohibit the use
of reagents, solutions, culture media,
control materials, calibration materials
and other supplies that have exceeded
their expiration date, have deteriorated or
are of substandard quality?
This should be stated in a policy or
procedure.
The laboratory must assign an
expiration date to any reagents
and media that do not have a
manufacturer-provided expiration date.
The assigned expiration date should
be based on known stability, frequency
of use, storage conditions and risk of
deterioration.
Physical observation during audit.
Requirement met?
Yes
No
Notes:
4.6
D5419
§493.1252(e)
Does the laboratory prohibit the
interchange of reagent kit components
of different lot numbers unless otherwise
specified by the manufacturer?
If applicable, this should be stated in a
policy or procedure.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Equipment Maintenance, Function Checks & Calibration | Page 23
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
5. ANALYTIC—EQUIPMENT MAINTENANCE, FUNCTION CHECKS & CALIBRATION
# SOM/CFR* Requirement Examples of Compliance Notes
5.1
D5429
§493.1254
(a)(1)
For unmodified manufacturer’s
equipment, instruments or test systems,
does the laboratory perform and
document maintenance as defined by
the manufacturer and with at least the
frequency specified by the manufacturer?
Current versions of equipment
manufacturer operator’s manuals
should be available (printed or
electronic).
Review manufacturer operator’s
manual to determine maintenance,
including frequency defined by the
manufacturer.
Review laboratory equipment
procedures, maintenance schedules
and records to ensure the lab
is performing and documenting
maintenance as defined by the
manufacturer.
Maintenance logs/preventative
maintenance (PM) schedule.
Documentation of repairs and service
requests.
Requirement met?
Yes
No
See:
ABI 7500 Semi-Annual Preventative
Maintenance Form
ABI 7500 Weekly and Monthly
Preventative Maintenance Form
Notes:
5.2
D5431
§493.1254
(a)(2)
For unmodified manufacturer’s equipment,
instruments or test systems, does the
laboratory perform and document function
checks as defined by the manufacturer
and with at least the frequency specified
by the manufacturer?
Note: Function checks must be within the
manufacturer’s established limits before
patient testing is conducted.
Current versions of equipment
manufacturer operator’s manuals
should be available (printed or
electronic).
Review manufacturer operator’s
manual to determine maintenance,
including frequency defined by the
manufacturer.
Review laboratory equipment
procedures, maintenance schedules
and records to ensure the lab
is performing and documenting
maintenance as defined by the
manufacturer.
QC records and logs.
Requirement met?
Yes
No
Notes:
5.3
D5433
§493.1254
(b)(1)
For equipment, instruments or test
systems developed in-house, modified
commercially available, or for which
maintenance and function checks are not
provided by the manufacturer, has the
laboratory:
Established a maintenance protocol
that ensures performance necessary for
accurate and reliable test results?
Performed and documented the
maintenance activities?
Procedure and/or maintenance schedule
defining maintenance requirements and
intervals for all equipment and records
used for diagnostic testing.
Note: This applies to equipment built in-
house, structurally-modified by the lab or
that is truly without a manufacturer-defined
PM.
See 5.1.
Requirement met?
Yes
No
See:
ABI 7500 Semi-Annual Preventative
Maintenance Form
ABI 7500 Weekly and Monthly
Preventative Maintenance Form
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Equipment Maintenance, Function Checks & Calibration | Page 24
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
5.4
D5435
§493.1254
(b)(2)
For equipment, instruments or test
systems developed in-house, modified
commercially available, or for which
maintenance and function checks are not
provided by the manufacturer, has the
laboratory:
Defined a function check protocol that
ensures performance that is necessary
for accurate and reliable test results
and test result reporting?
Does the laboratory perform and
document equipment, instrument and
test system function checks, including
background or baseline checks as
defined in 5.3?
Does the laboratory ensure that
function checks are within the
laboratory’s established limits before
patient testing is conducted?
Procedure and/or schedule defining
function check requirements for all
equipment and records of required
function check activities.
If function checks are out of range,
is there evidence that the laboratory
ensures corrected prior to patient
testing?
See 5.2.
Requirement met?
Yes
No
Notes:
5.5
D5437
§493.1255
(a)(1-3)
Does the laboratory perform and
document calibration procedures:
Following the manufacturer’s test
system instructions, using calibration
materials provided or specified,
and with at least the frequency
recommended by the manufacturer?
Using criteria verified or established by
the laboratory?
Using calibration materials appropriate
for the test system and, if possible,
traceable to a reference method or
reference material of known value
(including the number, type and
concentration of calibration materials,
as well as acceptable limits for and the
frequency of calibration)?
Whenever calibration verification fails to
meet the laboratory’s acceptable limits?
Generally applies to quantitative assays.
If applicable, records of calibration:
QC logs
Equipment maintenance logs
Run worksheets
SOPs, including procedure for
failures (QC failures) and supporting
documentation/follow up.
Review of laboratory test procedures to
meet manufacturer requirements.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Equipment Maintenance, Function Checks & Calibration | Page 25
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
5.6
D5439
§493.1255
(b)(1-3)
Does the laboratory perform and
document calibration verification
procedures:
Following the manufacturer’s calibration
verification instructions; using the
criteria verified or established by the
laboratory?
Including the number, type and
concentration of the materials,
as well as acceptable limits for
calibration verification?
Including at least a minimal (or zero),
mid-point and maximum value near
the upper limit of the range to verify
the laboratory’s reportable range?
Verifying calibration of test systems
at least once every six months and
whenever any of the following occur:
A complete change of reagents for a
procedure is introduced?
There is major preventive
maintenance or replacement of
critical parts?
Control materials reflect an unusual
trend or shift, or are outside the
laboratory’s acceptable limits?
The laboratory’s established
schedule for verifying the reportable
range for patient test results
requires more frequent calibration
verification?
Generally applies to quantitative assays;
review laboratory test procedures to
determine assays where this applies.
If applicable, records of calibration.
If an assay has a quantitative cut-off
value (e.g., some ELISAs), calibration
verification procedures may be needed
(contact the LD).
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Testing | Page 26
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
6. ANALYTIC—TESTING
6.1 Procedure Manual
# SOM/CFR* Requirement Examples of Compliance Notes
6.1.a
D5401
§493.1251 (a)
Does the laboratory have written
procedures for all tests, assays and
examinations performed available to
and followed by laboratory personnel?
A list of SOPs/test procedures/test
menus available for each CLIA-regulated
test performed may be stored and
accessed:
Paper-based
Electronic
Documentation for each test performed
must include:
The original start date the procedure
was used for CLIA-regulated testing, if
available
Approval by the technical supervisor
Approved by the current CLIA
laboratory director.
Requirement met?
Yes
No
Notes:
6.1.b
D5403
§493.1251
(b)(1-14)
Do procedure manuals include the
requirements laid out in 6.b.i-xiv below,
when applicable, for a test procedure?
SOP/test procedures must include the
required elements specified in 6.1.bi - xiv.
If these items are not found in the SOP,
they must be found elsewhere. May refer
to another procedure, such as:
Laboratory’s specimen submission
and handling procedure
Sample rejection policy, to include
information on what labeling is
accepted and not accepted
Sample collection instructions
Sample requisition form
Quality control
Corrective action/ non-conforming
event
Laboratory result reporting procedure
Requirement met?
Yes
No
Notes:
6.1.b.i
D5403
§493.1251
(b)(1)
Requirements for patient preparation;
specimen collection, labeling, storage,
preservation, transportation, processing
and referral; and criteria for specimen
acceptability and rejection as described
in §493.1242.
Requirement met?
Yes
No
Notes:
6.1.b.ii
D5403
§493.1251
(b)(2)
If microscopic examination is
performed, must include the detection
of inadequately prepared slides.
Requirement met?
Yes
No
Notes:
6.1.b.iii
D5403
§493.1251
(b)(3)
Step-by-step performance of the
procedure, including test calculations
and interpretation of results.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Testing | Page 27
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
6.1.b.iv
D5403
§493.1251
(b)(4)
Preparation of slides, solutions,
calibrators, controls, reagents, stains
and other materials used in testing.
Requirement met?
Yes
No
Notes:
6.1.b.v
D5403
§493.1251
(b)(5)
Calibration and calibration verification
procedures.
Requirement met?
Yes
No
Notes:
6.1.b.vi
D5403
§493.1251
(b)(6)
The reportable range for test results
for the test system as established or
verified in §493.1253.
Requirement met?
Yes
No
Notes:
6.1.b.
vii
D5403
§493.1251
(b)(7)
Control procedures.
Requirement met?
Yes
No
Notes:
6.1.b.
viii
D5403
§493.1251
(b)(8)
Corrective action to take when
calibration or control results fail to meet
the laboratory’s criteria for acceptability.
Requirement met?
Yes
No
Notes:
6.1.b.ix
D5403
§493.1251
(b)(9)
Limitations in the test methodology,
including interfering substances.
Requirement met?
Yes
No
Notes:
6.1.b.x
D5403
§493.1251
(b)(10)
Reference intervals (normal values).
Requirement met?
Yes
No
Notes:
6.1.b.xi
D5403
§493.1251
(b)(11)
Imminently life-threatening test results,
or panic or alert values.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Testing | Page 28
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
6.1.b.xii
D5403
§493.1251
(b)(12)
Pertinent literature references.
Requirement met?
Yes
No
Notes:
6.1.b.
xiii
D5403
§493.1251
(b)(13)
The laboratory’s system for entering
results in the patient record and
reporting patient results including, when
appropriate, the protocol for reporting
imminently life-threatening results, or
panic or alert values.
Requirement met?
Yes
No
Notes:
6.1.b.
xiv
D5403
§493.1251
(b)(14)
Description of the course of action
to take if a test system becomes
inoperable.
Requirement met?
Yes
No
Notes:
6.1.c
D5405
§493.1251
(c)
If the laboratory uses manufacturer’s
test systems or operator’s manuals
(e.g., LRN or EUA), does the laboratory
provide supplemental information to
meet requirements of §493.1251
(b)(1-14)?
Supplemental documention may include:
Laboratory-specific specimen
acceptance/rejection criteria
Reference (normal) values
Alert values
Requirement met?
Yes
No
Notes:
6.1.d
D5407
§493.1251
(d)
Are procedures and changes to
procedures approved, signed and dated
by the current laboratory director before
use?
Documentation of CLIA laboratory
director approval for CLIA regulated test
procedures.
Requirement met?
Yes
No
Notes:
6.1.e
D6106
§493.1445
(e)(14)
Technical Supervisor Responsibility
(Delegated by LD): Does the technical
supervisor ensure that an approved
procedure manual is available to all
personnel responsi-ble for any aspect of
the testing process?
See evidence above for 6.1.a.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
6.1.f
D5409
§493.1251 (e)
D3029
§493.1105
(a)(2)
Record Retention: Is a copy of test
procedures retained for at least two
years after the procedure has been
discontinued and does the procedure
include dates of initial use and
discontinuance?
Note: Check state requirements as
applicable—go with most stringent.
Are discontinued SOP/test procedures
available if requested and include the
dates of initial use and date testing was
discontinued?
SOPs or testing procedures must be
available from the past two years,
including ones that have been
discontinued.
Longer retention times might be in
place for other state, local and federal
regulations and must also be followed.
The CLIA two-year rule must be followed
at a minimum.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Testing | Page 29
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
6.2 Analytic Systems—Verification and Validation (establishment of performance
specifications)
# SOM/CFR* Requirement Examples of Compliance Notes
6.2.a
D5419
§493.1253 (a)
Has the laboratory verified or established
performance specifications (validated)
all tests put into place or modified on or
after April 24, 2003 prior to reporting
patient results?
Note: Verification or validation for tests
used prior to April 24, 2003 is highly
recommended.
There is an SOP that includes
verification and validation procedure.
Validation or verification records for
each test.
Requirement met?
Yes
No
Notes:
6.2.b
D6085
§493.1445
(e)(3)(i)
Technical Supervisor Responsibility
(Delegated by LD): Does the
technical supervisor ensure that test
methodologies selected have the
capability of providing the quality of
results required for patient care?
See evidence for 6.2.c and 6.2.e.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
6.2.c
D5421
§493.1253
(b)(1)
For unmodified FDA-cleared/approved
test systems, has the laboratory verified
the following prior to reporting patient
results:
The lab can obtain performance
specifications comparable to those
established by the manufacturer for
accu-racy, precision and reportable
range?
The manufacturer’s reference intervals
(normal values) are appropriate for
their population?
Documentation and records of
verification/validation must be
available for inspection.
Records must include evidence of
review and approval by the technical
supervisor.
Requirement met?
Yes
No
Notes:
6.2.d
D6086
§493.1445
(e)(3)(ii)
Technical Supervisor Responsibility
(Delegated by LD): Does the technical
supervisor ensure that verification
procedures used are adequate to
determine the accuracy, precision
and other pertinent performance
characteristics of the method?
See evidence for 6.2.c and 6.2.e.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Testing | Page 30
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
6.2.e
D5423
§493.1253
(b)(2)
For all modified FDA-cleared/approved
test systems or a laboratory developed
test system in which the performance
specifications are not provided by
the manufacturer, has the laboratory
established the following performance
characteristics:
Accuracy?
Precision?
Analytical sensitivity?
Analytical specificity to include
interfering substances?
Reportable range of test results?
Reference intervals (normal values)?
Any other performance characteristic
required for test performance?
Documentation and records of
verification/validation must be
available for inspection.
Records must include evidence of
review and approval by the technical
supervisor.
Requirement met?
Yes
No
Notes:
6.2.f
D6095
§493.1445
(e)(6)
Technical Supervisor Responsibility
(Delegated by LD): Does the technical
supervisor ensure the establishment
and maintenance of acceptable levels
of analytical perfor-mance for each test
system?
New or modified test procedure
verification/validation studies are
submitted to the LD for review and
approval
See evidence for 6.2.c and 6.2.e.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
6.2.g
D5425
§493.1253
(b)(3)
Has the laboratory determined each
test system’s calibration and control
procedures based upon the performance
specifications verified or established
under §493.1253(b)(1) or §493.1253(b)
(2)?
Determination of calibration, if
applicable, and control requirements
must be in validation records
Test procedures
Requirement met?
Yes
No
Notes:
6.2.h
D5427
§493.1253
(c)
Does the laboratory document all
activities for all method verification
and establishment of performance
specifications?
The laboratory must have documentation
and records available for inspection for
verification/validation activities including
records of the actual measurements
taken, reactions and/or observations.
Requirement met?
Yes
No
Notes:
6.2.i
D3033
§493.1105
(a)(3)(i)
Record Retention: Are records of test
system performance specifications that
the laboratory establishes or verifies
under §493.1253 retained for the period
of time the laboratory uses the test
system but no less than two years?
Note: Longer retention times might be
in place for other state, local, federal
regulations and must also be followed. The
CLIA two-year rule must be followed at a
minimum.
Verfication/validation procedures must
be available for the period of the time
that the laboratory uses the testing
system, but no less than two years.
Check state requirements as applicable—
go with most stringent.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Quality Control | Page 31
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
7. ANALYTIC—QUALITY CONTROL
7.1 QC—General
# SOM/CFR* Requirement Examples of Compliance Notes
7.1.a
D5441
§493.1256
(a)(b)(c)(1-2)
Has the laboratory established quality
control (QC) procedures for each test
system that monitor the accuracy and
precision of the complete analytic
process?
Does the QC procedure:
Detect immediate errors that occur
due to test system failure, adverse
environmental conditions and operator
performance?
Monitor over time the accuracy
and precision of test performance
that may be influenced by changes
in test system performance and
environmental conditions, and
variance in operator performance?
Comply with the recommendations of
the manufacturer?
Does the QC procedure establish the
number, type and frequency of testing
control materials using, if applicable, the
performance specifications verified or
established by the laboratory?
Quality Control policy/procedure
Records
Logs or worksheets
Note: Specifics for number, type and
frequency needs to be defined in each
individual test procedure.
Requirement met?
Yes
No
Notes:
7.1.b
D6117
§493.1451
(b)(4)
Technical Supervisor Responsibility: Has
the technical supervisor established a
quali-ty control program appropriate for
the testing performed covering the entire
path of work flow.
QC policies and procedures.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
7.1.c
D6093
§493.1445
(e)(5)
Technical Supervisor Responsibility
(Delegated by LD): Does the technical
supervisor ensure that the quality control
and quality assessment programs are
established and maintained to assure the
quality of laboratory services provided
and to identify fail-ures in quality as they
occur?
QC policies/procedures and records of
review.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
7.1.d
D6148
§493.1463
(a)(4)
General Supervisor Responsibility:
Does the general supervisor monitor test
analyses and specimen examinations to
ensure that acceptable levels of analytic
performance are maintained?
QC is reviewed and failed QC is
documented and investigated.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
General Supervisor Personnel Form
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Quality Control | Page 32
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
7.1.e
D5445
§493.1256
(d)(1-2)
Does the laboratory perform control
procedures as:
Defined in this section and/or as
specified in the additional specialty
and subspecialty requirements,
Recommended by the manufacturer
or established by the laboratory
ONLY when they meet or exceed the
requirement in this section, or
Developed and customized by the
laboratory using an Individualized
Quality Control Plan (IQCP) to meet
requirements, with IQCP records
reviewed and verified annually as
appropriate to the test methods.
QC policies/procedures
QC records, logs or worksheets
includes lot number, date received,
opened or prepared date and
expiration date.
IQCP documentation reviewed annually
for potential new risks and adequacy
of the QC plan.
Actual measurements, reactions and/
or observations should be recorded.
Evaluation of QC failures and, if
applicable, nonconforming event and
corrective actions records that include:
Retrospective review of patient
results from testing since the last
time the system was known to be
performing within specification
Issuing corrected patient results or
contacting physicians to notify of
potential error.
Evidence of supervisory review.
Check that patient results were not
reported when QC was unacceptable.
Requirement met?
Yes
No
See Quality Assessment Worksheet
Notes:
7.1.f
§493.1256
(d)(3)
Does the laboratory perform QC
procedures at least once each day
patient speci-mens are assayed for the
following (7.1f.i-v):
See evidence for 7.1.e.
Check QC records to ensure appropriate
controls are performed for each test
procedure.
Requirement met?
Yes
No
Notes:
7.1.f.i
D5447
§493.1256
(d)(3)(i)
For quantitative procedures, include
two control materials of different
concentrations.
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.1.f.ii
D5449
§493.1256
(d)(3)(ii)
For qualitative procedures, include a
negative and positive control material.
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.1.f.iii
D5451
§493.1256
(d)(3)(iii)
For test procedures producing a graded
or titered result, include a negative
control material and a control material
with graded or titered reactivity.
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Quality Control | Page 33
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
7.1.f.iv
D5453
§493.1256
(d)(3)(iv)
For test systems that have an extraction
phase, include two control materials,
including one that is capable of detecting
errors in the extraction process.
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.1.f.v
D5453
§493.1256
(d)(3)(v)
For molecular amplification procedures,
include two controls and, if reaction
inhibition is a significant source of false
negative results, a control material
capable of detect-ing the inhibition?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.1.g
D5457
§493.1256
(d)(4)
For thin layer chromatography (TLC),
spot each plate or card with a calibrator
containing all known substances or
drug groups, as appropriate which are
identified by TLC and reported by the
laboratory; and include at least one
control material on each plate or card
which is processed through each step of
patient testing, including extrac-tion?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.1.h
D5459
§493.1256
(d)(5)
For electrophoresis procedures, include,
concurrent with patient specimens, at
least one control material containing
the substances being identified or
measured?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.1.i
D5461
§493.1256
(d)(6)
Does the laboratory perform specified
quality control before resuming patient
testing when a:
Complete change of reagents is
introduced?
Major preventive maintenance is
performed or a critical part that
may influence test performance is
replaced?
Quality control policy/procedure
Records
Logs or worksheets
IQCP documentation
Instrument logs
Non-conforming event (NCE) log
Reagent log
Requirement met?
Yes
No
Notes:
7.1.j
D5463
§493.1256
(d)(7)
Does the laboratory rotate control
material testing among all personnel who
perform patient testing?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.1.k
D5465
§493.1256
(d)(8)
Does the laboratory test control
materials in the same manner as patient
specimens?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Quality Control | Page 34
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
7.1.l
D5467
§493.1256
(d)(9)
For quantitative procedures, when using
calibration material as a control material,
are different lot numbers used from the
calibration material used to calibrate the
test system?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.1.m
D5459
§493.1256
(d)(10)
For quantitative procedures, has the
laboratory established or verified the
criteria for acceptability of all control
materials?
Are statistical parameters (e.g., mean
and standard deviation) defined for
each lot of control materials from
which quantitative results are derived?
If the laboratory uses commercially
assayed control material values, are
the method and instrumentation val-
ues verified by the laboratory?
If the laboratory uses unassayed
control material, are statistical
parameters established over time
through con-current testing of
control materials having previously
determined statistical parameters?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.1.n
D5471,
D5473,
D5475
§493.1256
(e)(1-3)
For reagent, media and supply checks,
does the laboratory perform and
document the following:
Check each batch (prepared in-
house), lot number (commercially
prepared) and shipment of reagents,
disks, stains, antisera (excluding
bacteriology) and identification
systems (using two or more substrates
or reagents) when prepared or opened
for positive and negative reactivity, as
well as graded reactivity, if applicable?
Each day of use, test staining material
for intended reactivity? Controls are
to include both positive and nega-tive
reactivity.
Each time of use, check fluorescent
and immunohistochemical stains for
positive and negative reactivity?
See evidence for 7.1.e
See media logs
Requirement met?
Yes
No
Notes:
7.1.o
D5477
§493.1256
(e)(4)
Before or concurrent with initial use
media, does the laboratory check and
document each batch for:
Sterility, if sterility is required for
testing?
Ability to support growth and, as
applicable, select or inhibit specific
organisms or produce biochemical
responses?
Physical characteristics of the media
when compromised and report any
deterioration in the media to the
manufacturer?
See evidence for 7.1.e
See media logs
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Quality Control | Page 35
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
7.1.p
D5479
§493.1256
(e)(5)
Does the laboratory follow the
manufacturer’s specifications for using
reagents, media and supplies?
See evidence for 7.1.e
Compare to manufacturer’s package
insert.
Requirement met?
Yes
No
Notes:
7.1.q
D5481
§493.1256
(f)(g)
Does the laboratory document quality
control testing and evaluate control
results to ensure established criteria for
acceptability are met before reporting
patient test results?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.1.r
D6118
§493.1451
(b)(5)
Technical Supervisor Responsibility:
Does the technical supervisor ensure
that technical problems are resolved
and that remedial actions are taken
and documented when laboratory
test systems deviate from established
performance specifications?
QC policies and procedures documented;
if delegated in writing to the GS (see
7.1.t).
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
7.1.s
D6119
§493.1451
(b)(6)
Technical Supervisor Responsibility:
Does the technical supervisor ensure
that patient test results are not reported
until corrective actions have been
taken and test sys-tems are functioning
properly?
QC policies and procedures documented;
if delegated in writing to the GS (see
7.1.t).
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
7.1.t
D6149,
D6150
§493.1463
(b)(1-2)
General Supervisor Responsibility
(If delegated by LD or TS): Does the
general supervisor assure all remedial
actions are taken whenever test systems
deviate from the la-boratory’s established
performance specifications and ensure
patient test results are not reported
until all corrective actions have been
taken and the test system is properly
functioning?
Failed QC is documented and
investigated.
NCE corrective action followed up for
effectiveness, etc.
Patient results are not reported when
QC out of range.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
General Supervisor Personnel Form
Notes:
7.1.u
D6096,
D6097
§493.1445
(e)(7)
Technical Supervisor Responsibility
(Delegated by LD): Does the technical
supervisor ensure that all necessary
remedial actions are taken and
documented whenever significant
deviations from the laboratory’s
established performance characteristics
are identified, and that patient test
results reported only when the system is
functioning properly?
QC policies and procedures documented;
if delegated in writing to the GS (see
7.1.t).
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Quality Control | Page 36
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
7.1.v
D3031
§493.1105
(a)(3)
Record Retention: Are quality control
and patient test records (including
instrument printouts, if applicable)
and records documenting all analytic
activities (specified in §493.1252
through 493.1289) retained for at least
two years?
Note: Longer retention times might be in
place for other state, local and federal
regulations and must also be followed. The
CLIA two-year rule must be followed at a
minimum.
QC records must be available from the
past two years, including ones that have
been discontinued.
The records must include instrument
charts, graphs, printouts, transcribed
data and manufacturers’ assay
information sheets for control and
calibration materials. If data are
transcribed, ensure that the original and
the transcribed copy are retained for
two years. Printouts from an instrument
that is not directly interfaced with the
laboratory information system must be
retained for two years.
Note: Check state requirements as
applicable—go with most stringent.
Requirement met?
Yes
No
Notes:
7.2 QC—Bacteriology
# SOM/CFR* Requirement Examples of Compliance Notes
7.2.a
D5501, D5503,
D5505
§493.1261
(a)(1-3)
Does the laboratory check the following
for positive and negative reactivity using
control organisms:
Each day of use for beta lactamase
methods other than Cefinase
TM
?
Each week of use for gram stains?
When each batch (prepared in house),
lot number (commercially prepared)
and shipment of antisera is pre-pared
or opened, and once every six months
thereafter?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.2.b
D5507
§493.1261
(b)(1-2)(c)
For antimicrobial susceptibility tests,
does the laboratory:
Check each batch of media and
each lot number and shipment of
antimicrobial agent(s) before, or
concurrent with initial use, using
approved control organisms?
Test approved control organisms each
day of testing?
Document and evaluate zone sizes or
MICs for control organisms to ensure
that they are within established limits
before reporting patient results?
Document all control procedures
performed?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Quality Control | Page 37
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
7.3 QC—Mycobacteriology
# SOM/CFR* Requirement Examples of Compliance Notes
7.3.a
D5511
§493.1262 (a)
On each day of use, does the laboratory
check all reagents or test procedures
used for mycobacteria identification
with at least one acid-fast organism that
produces a positive reaction and an acid-
fast organism that produces a negative
reaction?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.3.b
D5513
§493.1262
(b)(1-3)(c)
For antimycobacterial susceptibility
tests, does the laboratory check each
batch of media and each lot number and
shipment of antimycobacterial agent(s)
before, or concurrent with, initial use
using appropriate control organism(s)?
If so,
Has the laboratory established limits
for acceptable control results?
Does the laboratory use appropriate
control organism(s) each week
tests are performed to check the
procedure?
Are the results of control organism(s)
verified to be within established limits
before patient results are reported?
Does the laboratory document all
control procedures performed?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.4 QC—Mycology
# SOM/CFR* Requirement Examples of Compliance Notes
7.4.a
D5517
§493.1263 (a)
Does the laboratory check each batch
(prepared in-house), lot number
(commercially prepared) and shipment of
lactophenol cotton blue when prepared
or opened for intended reactivity with a
control organism(s)?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.4.b
D5519
§493.1263 (b)
(1-3)(c)
For antifungal susceptibility tests, does
the laboratory check each batch of media
and each lot number and shipment of
antifungal agent(s) before, or concurrent
with, initial use using appropriate control
organism(s)?
If so,
Has the laboratory established limits
for acceptable control results?
Does the laboratory use appropriate
control organism(s) each week
tests are performed to check the
procedure?
Are the results of control organism(s)
verified to be within established limits
before patient results are reported?
Does the laboratory document all
control procedures performed?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Analytic—Quality Control | Page 38
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
7.5 QC—Parasitology
# SOM/CFR* Requirement Examples of Compliance Notes
7.5.a
D5523
§493.1264 (a)
Does the laboratory have a reference
collection of slides or photographs
available and, if available, gross
specimens for identification of parasites
and use these references in the
laboratory for appropriate comparison
with diagnostic specimens?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.5.b
D5525
§493.1264 (b)
Does the laboratory calibrate and use
ocular micrometers for determining
the size of ova and parasites, if size is
critical?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.5.c
D5527
§493.1264
(c)(d)
Does the laboratory check permanent
stains each month of use using a
fecal sample control material that will
demonstrate staining characteristics?
Does the laboratory document all control
procedures performed?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.6 QC—Virology
# SOM/CFR* Requirement Examples of Compliance Notes
7.6.a
D5531
§493.1265
(a)(b)
When using cell cultures to isolate or
identify viruses, does the laboratory
simultaneously incubate a cell substrate
control or uninoculated cells as a
negative control?
Does the laboratory document all control
procedures performed?
See evidence for 7.1.e.
Requirement met?
Yes
No
Notes:
7.7 QC—Other than Microbiology
# SOM/CFR* Requirement Examples of Compliance Notes
7.7.a
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | PT & Alternative Assessment | Page 39
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
8. PROFICIENCY TESTING (PT) & ALTERNATIVE ASSESSMENT
8.1 PT—Enrollment
For all requirements below, refer to the Proficiency Testing and Alternative Assessment Review Worksheet Tool.
# SOM/CFR* Requirement Examples of Compliance Notes
8.1.a
D2001
§493.801 (a)
Is the laboratory enrolled in a CMS-
approved PT program offering three
(3) events per year (two (2) events for
mycobacteriology) with a minimum of five
(5) samples per event for all regulated
analytes tested?
Compare test lists to PT enrollment
Records of participation in an
approved PT program for labs testing
regulated analytes
Requirement met?
Yes
No
Notes:
8.1.b
D6088
§493.1445
(e)(4)
Technical Supervisor Responsibility
(Delegated by LD): Does the technical
supervisor ensure that the laboratory is
enrolled in an HHS-approved proficiency
testing program, if available, for the
testing performed?
See evidence for 8.1.a
Note: Duplicate of 8.1.c if delegated in
writing by the LD.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
8.1.c
D6116
§493.1451
(b)(3)
Technical Supervisor Responsibility: Does
the technical supervisor ensure that the
laboratory is enrolled and participates
in an HHS approved proficiency testing
program commensurate with the services
offered?
See evidence for 8.1.a.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
8.1.d
D5217, D5219
§493.1236
(c)(1-2)
Does the laboratory verify the accuracy at
least twice annually using an alternative
assessment procedure (AAP) such as
other external assessment programs,
blind testing, or split samples for other
testing such as:
Tests where no compatible PT samples
are offered by a CMS approved PT
program?
Tests for nonregulated analytes?
Compare test lists to what is not
covered in PT enrollment and must
perform an AAP
Procedure and records indicating
at least twice annual alternative
assessment for all other testing
Records should be easy to follow and
should include a summary page with
expected results, actual results and
evaluation of performance
Requirement met?
Yes
No
Notes:
8.2 PT—Testing of Samples
For all requirements below, refer to the Proficiency Testing and Alternative Assessment Review Worksheet Tool.
# SOM/CFR* Requirement Examples of Compliance Notes
8.2.a
D2006
§493.801 (b)
Does the laboratory examine PT samples
in the same manner as it tests patient
specimens?
Policy/procedure to ensure
compliance
QA manual
Attestation statements provided by
CAP, other PT provider or created
in-house
Requirement met?
Yes
No
See the Audit Attestation Form
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | PT & Alternative Assessment | Page 40
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
8.2.b
D6089
§493.1445
(e)(4)(i)
Technical Supervisor Responsibility
(Delegated by LD): Does the technical
supervisor ensure that proficiency
testing samples are tested as required by
§493.801?
See evidence for 8.2.a.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
8.2.c
D6176
§493.1495
(b)(2)
Testing Personnel Responsibility: Do
testing personnel maintain records that
demonstrate that proficiency testing
samples are tested in the same manner
as patient specimens?
Worksheets and sample logs
Policies and procedures
QA manual
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Testing Personnel Form
Notes:
8.2.d
D2007
§493.801
(b)(1)
Are PT samples examined along with the
lab’s regular workload by personnel who
routinely perform testing using the lab’s
routine methods?
PT records indicate that testing was
performed by testing personnel
authorized to perform the assay
Attestation statements
PT records to confirm testing
personnel are rotated
Requirement met?
Yes
No
Notes:
8.2.e
D2009
§493.801
(b)(1)
Does the individual testing the PT
samples and the technical supervisor
(delegated by the LD) sign an attestation
statement that PT samples are integrated
into the patient workload using the
laboratory’s routine methods?
Attestation statements provided by
the College of American Pathologists,
other PT provider or created in-house
Policies and procedures
Requirement met?
Yes
No
Notes:
8.2.f
D2010
§493.801
(b)(2)
Are PT samples tested the same number
of times as patient samples are routinely
tested?
Worksheets and sample logs
Policies and procedures
Requirement met?
Yes
No
Notes:
8.2.g
D2011, D2013
§493.801
(b)(3-4)
Does the laboratory have procedures to
ensure:
No inter-laboratory communication
regarding PT samples/results occurs
until after the result submission
deadline date has passed?
PT samples are not referred to
another laboratory for analysis, or if
PT samples are received from another
laboratory for testing, the samples are
not tested and CMS is notified?
Policies and procedures
QA manual
Worksheets, sample logs and final
reports for verification
Requirement met?
Yes
No
Notes:
8.2.h
D2015
§493.801
(b)(5)
Does the laboratory document the
handling, preparation, processing,
examination, and each step in the
testing and reporting of results for all PT
samples?
Policies and procedures
QA manual
Worksheets, sample logs and final
reports for verification
Attestation statements
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | PT & Alternative Assessment | Page 41
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
8.3 PT—Evaluation of Performance
For all requirements below, refer to the Proficiency Testing and Alternative Assessment Review Worksheet Tool.
# SOM/CFR* Requirement Examples of Compliance Notes
8.3.a
D6090
§493.1445
(e)(4)(ii)
Technical Supervisor Responsibility
(Delegated by LD): Does the technical
supervisor ensure that PT results
are returned within the timeframes
established by the PT program?
PT records are reviewed and complete.
Requirement met?
Yes
No
Notes:
8.3.b
D5211
§493.1236 (a)
Does the laboratory review and evaluate
the results obtained on PT?
Has the laboratory had unsatisfactory PT
responses? Records should indicate:
PT result evaluations are reviewed and
signed by the technical supervisor
Documentation of investigation of
the incorrect result, including any
corrective action or retraining required
For assays where failed results are
common, results should be logged
for trends that might likely cause the
failed results
NCEs, self-assessments, corrective
actions
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
8.3.c
D6091
§493.1445
(e)(4)(iii)
Technical Supervisor Responsibility
(Delegated by LD): Does the technical
supervisor ensure that all PT reports
received are reviewed by the appropriate
staff to evaluate the laboratory’s
performance and to identify any problems
that require corrective actions?
PT performance is reviewed with
appropriate staff.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
8.3.d
D5213–D5215
§493.1236 (b)
Does the laboratory verify the accuracy of
any PT result that was not scored for any
reason?
Records of review and evaluation of
ungraded PT challenges.
Requirement met?
Yes
No
Notes:
8.3.e
D5221
§493.1236 (d)
Are PT evaluations and verification
activities documented?
Note: This applies even if the overall PT
event was successful.
Records available.
Requirement met?
Yes
No
Notes:
8.3.f
D6092
§493.1445
(e)(4)(iv)
Technical Supervisor Responsibility
(Delegated by LD): Does the technical
supervisor ensure that an approved
corrective actions plan is followed when
any proficiency testing result is found to
be unacceptable or unsatisfactory?
Documentation of review/approval.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | PT & Alternative Assessment | Page 42
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
8.3.g
D3037
§493.1105
(a)(4)
Record Retention: Are records of PT
examination and documentation of
activities retained for a minimum of two
years after the PT event?
Note: Longer retention times might be in
place for other state, local and federal
regulations and must also be followed. The
CLIA two-year rule must be followed at a
minimum.
Obtain signed attestation statements,
PT results and scores from the provider,
documentation of review and records of
any corrective actions, either electronic
or hard copies.
Note: Check state requirements as
applicable—go with most stringent.
Requirement met?
Yes
No
Notes:
8.4 Comparison of Test Results
# SOM/CFR* Requirement Examples of Compliance Notes
8.4.a
D5775
§493.1281
(a)(c)
If the laboratory performs the same
test using different methodologies,
instruments or testing sites, does
the laboratory have a procedure that
twice a year evaluates and defines the
relationship between test results using
the different methodologies, instruments
or testing sites?
Does the laboratory document test result
comparison?
Procedure and records for comparison of
results when the same test is:
Performed by another testing site
Performed using multiple instruments
Performed by a different methodology.
The laboratory must have written criteria
for acceptable differences in test values.
This comparison must be done twice
a year and must be approved by the
technical supervisor.
Requirement met?
Yes
No
See examples of instrument comparison
records:
EVOLIS Instrument Comparability Chart
PCR Instrument Comparison Example
Notes:
8.4.b
D5777
§493.1281
(a)(c)
Does the laboratory have a system to
assess patient test results that appear
inconsistent with the following relevant
criteria, when available:
Patient Age?
Patient Sex?
Diagnosis or pertinent clinical data?
Distribution of patient test results?
Relationship with other test
parameters?
If applicable, documentation within a
test SOP.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | General Laboratory Systems & Retention Requirements | Page 43
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
9. GENERAL LABORATORY SYSTEMS & RETENTION REQUIREMENTS
9.1 General Laboratory Systems §493.1230
# SOM/CFR* Requirement Examples of Compliance Notes
9.1.a
D5787, D5789
§493.1283
(a)(1-4)(b)
Does the laboratory maintain an
information or record system that
includes the following:
Positive identification of the
specimen?
Date and time of specimen receipt into
the laboratory?
The condition and disposition of
specimens that do not meet the
laboratory’s criteria for specimen
acceptability?
Records and dates of specimen
testing, including the identity of the
personnel who performed the test(s)?
Records of patient testing including, if
applicable, instrument printouts?
Unique identifiers to provide positive
identification of specimens from
accessioning through storage.
Receipt date and time is recorded for
all specimens.
Condition of specimens and
disposition is recorded.
Verify procedure for positive specimen
identification, log date, time of receipt
and log rejected specimens.
Worksheets should indicate date and
time of testing and include the identity
of the individual performing the test
(initials, signature or electronically
logged).
Instruments printouts maintained, if
applicable.
Requirement met?
Yes
No
Notes:
9.1.b
D5201
§493.1231
Does the laboratory ensure
confidentiality of patient information
throughout all phases of the total testing
processes that are under its control?
Controlled access of laboratory
areas where patient info may easily
be viewed in a hard copy format or
electronically.
Required security awareness training
for all employees.
Observe during walkthrough if patient
reports are lying out in open view.
Staff access permissions forms, if
applicable.
Requirement met?
Yes
No
Notes:
9.1.c
D5203
§493.1232
Has the laboratory established and does
it follow written policies and procedures
that ensure positive identification and
optimum integrity of patient samples
from time of receipt through completion
of testing and reporting of results?
Policy/procedure to ensure
compliance.
Unique identifiers to provide positive
identification of specimens from
accessioning through storage.
Observe during walkthrough how
aliquots are labeled on the bench.
Requirement met?
Yes
No
Notes:
9.1.d
D5205
§493.1233
Does the laboratory have a system in
place to document all complaints and
problems reported to the laboratory
and does the laboratory conduct and
document investigations of complaints,
when appropriate?
Log of communications and
complaints including follow up and
date closed, reviewed for trends of
problems and corrective action taken
(NCEs)
Policy and procedure on how to define
and address complaints.
Requirement met?
Yes
No
Notes:
9.1.e
D5207
§493.1234
Does the laboratory have a system in
place to identify and document problems
that occur as a result of a breakdown in
communication between the laboratory
and an authorized person who orders or
receives test results?
Review communication logs, NCE logs
and reports for problems that may be
related to communication.
Example: Ongoing inquiries from submitters
for clarification concerning appropriate
specimen, proper collection, transport, etc.
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Quality Assessment & Corrective Action | Page 44
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
10. QUALITY ASSESSMENT & CORRECTIVE ACTION
10.1 Quality Assessment
# SOM/CFR* Requirement Examples of Compliance Notes
10.1.a
D5291
§493.1239 (a)
General Laboratory Systems
Has the laboratory established and
does it follow written policies and
procedures for an ongoing quality
assessment mechanism to monitor,
assess, continuously improve and, when
indicated, correct problems identified in
the general laboratory systems?
SOPs, quality metrics, internal audits,
non-conforming events, corrective action
reports.
QA of general lab systems includes
assessing practices/issues related to:
Patient confidentiality
Specimen identification and integrity
Complaint investigations
Communications
Personnel Competency
Proficiency testing performance
Requirement met?
Yes
No
Notes:
10.1.b
D5391
§493.1249 (a)
Pre-analytic Systems
Has the laboratory established and
does it follow written policies and
procedures for an ongoing quality
assessment mechanism to monitor,
assess, continuously improve and, when
indicated, correct problems identified in
the preanalytic systems?
SOPs, quality metrics, internal audits,
non-conforming events, corrective action
reports.
QA of pre-analytic lab systems includes
assessing practices/issues related to:
Test requests
Specimen submission
Specimen handling
Tracking of rejected specimens and
reasons why
Verification of requisitions
Patient identifiers
Does the laboratory track problems with
incorrect or incomplete information on
test requisitions?
Requirement met?
Yes
No
Notes:
10.1.c
D5791
§493.1289 (a)
Analytic Systems
Has the laboratory established and
does it follow written policies and
procedures for an ongoing quality
assessment mechanism to monitor,
assess, continuously improve and, when
indicated, correct problems identified in
the analytic systems?
SOPs, quality metrics, internal audits,
non-conforming events, corrective action
reports.
QA of analytic lab systems includes
assessing practices/issues related to:
Test procedures
Accurate and reliable test systems,
equipment, instruments, reagents,
materials and supplies
Specimen and reagent storage
conditions
Equipment/instrument/test/system
maintenance and function checks
Establishment and verification of
method performance specifications
Calibration and calibration verification
Control procedures
Semiannual comparison of tests
results for the same analyte
Corrective actions
Test records
Patient identifiers
Requirement met?
Yes
No
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Quality Assessment & Corrective Action | Page 45
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
10.1.d
D5891
§493.1299 (a)
Post-analytic Systems
Has the laboratory established and
does it follow written policies and
procedures for an ongoing quality
assessment mechanism to monitor,
assess, continuously improve and, when
indicated, correct problems identified in
the post-analytic systems?
SOPs, quality metrics, internal audits,
non-conforming events, corrective action
reports.
QA of post-analytic lab systems includes
assessing practices/issues related to:
Accuracy and completeness
Turn-around time monitoring and
notification
Notification for abnormal or panic
values
Tracking corrected reports
Patient identifiers
Requirement met?
Yes
No
Notes:
10.1.e
D5293
§493.1239 (b)
General Laboratory Systems
Does the general laboratory systems
quality assessment include a review of
the following:
Effectiveness of corrective actions
taken to resolve problems?
Effectiveness of revisions to
policies and procedures to prevent
recurrence of problems?
Discussion of general laboratory
systems quality assessment reviews
with appropriate staff?
Management review or minutes from
quality management meetings with
staff.
Quality management reports.
Corrective action reports include
actions to prevent recurrence
of problem and plan to monitor
effectiveness.
Requirement met?
Yes
No
Refer to 10.1.a-d.
Notes:
10.1.f
D5393
§493.1249 (b)
Pre-analytic Systems
Does the pre-analytic systems quality
assessment include a review of the
following:
Effectiveness of corrective actions
taken to resolve problems?
Effectiveness of revisions to
policies and procedures to prevent
recurrence of problems?
Discussion of general laboratory
systems quality assessment reviews
with appropriate staff?
Management review or minutes from
quality management meetings with
staff.
Quality management reports.
Corrective action reports include
actions to prevent recurrence
of problem and plan to monitor
effectiveness.
Requirement met?
Yes
No
Refer to 10.1.a-d.
Notes:
10.1.g
D5793
§493.1289 (b)
Analytic Systems
Does the analytic systems quality
assessment include a review of the
following:
Effectiveness of corrective actions
taken to resolve problems?
Effectiveness of revisions to
policies and procedures to prevent
recurrence of problems?
Discussion of general laboratory
systems quality assessment reviews
with appropriate staff?
Management review or minutes from
quality management meetings with
staff.
Quality management reports.
Corrective action reports include
actions to prevent recurrence
of problem and plan to monitor
effectiveness.
Requirement met?
Yes
No
Refer to 10.1.a-d.
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Quality Assessment & Corrective Action | Page 46
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
10.1.h
D5893
§493.1299 (b)
Post-analytic Systems
Does the post-analytic systems quality
assessment include a review of the
following:
Effectiveness of corrective actions
taken to resolve problems?
Effectiveness of revisions to
policies and procedures to prevent
recurrence of problems?
Discussion of general laboratory
systems quality assessment reviews
with appropriate staff?
Management review or minutes from
quality management meetings with
staff.
Quality management reports.
Corrective action reports include
actions to prevent recurrence
of problem and plan to monitor
effectiveness.
Requirement met?
Yes
No
Refer to 10.1.a-d.
Notes:
10.1.i
D5293
§493.1239 (c)
General Laboratory Systems
Does the laboratory document all
quality assessment activities for general
laboratory systems?
Documented evidence includes:
Records generated per SOPs related to
QA activities
Documentation of quality metrics
NCEs
Corrective action reports
Management review presentations
or minutes from quality management
meetings with staff
Quality management reports.
Records must show evidence of
management or supervisory review.
Requirement met?
Yes
No
Refer to 10.1.a-d.
Notes:
10.1.j
D5393
§493.1249 (c)
Pre-analytic Systems
Does the laboratory document all
quality assessment activities for pre-
analytic systems?
Documented evidence includes:
Records generated per SOPs related to
QA activities
Documentation of quality metrics
NCEs
Corrective action reports
Management review presentations
or minutes from quality management
meetings with staff
Quality management reports.
Records must show evidence of
management or supervisory review.
Requirement met?
Yes
No
Refer to 10.1.a-d.
Notes:
10.1.k
D5793
§493.1289 (c)
Analytic Systems
Does the laboratory document all
quality assessment activities for analytic
systems?
Documented evidence includes:
Records generated per SOPs related to
QA activities
Documentation of quality metrics
NCEs
Corrective action reports
Management review presentations
or minutes from quality management
meetings with staff
Quality management reports.
Records must show evidence of
management or supervisory review.
Requirement met?
Yes
No
Refer to 10.1.a-d.
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Quality Assessment & Corrective Action | Page 47
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
# SOM/CFR* Requirement Examples of Compliance Notes
10.1.l
D5893
§493.1299 (c)
Post-analytic Systems
Does the laboratory document all
quality assessment activities for post-
analytic systems?
Documented evidence includes:
Records generated per SOPs related to
QA activities
Documentation of quality metrics
NCEs
Corrective action reports
Management review presentations
or minutes from quality management
meetings with staff
Quality management reports.
Records must show evidence of
management or supervisory review.
Requirement met?
Yes
No
Refer to 10.1.a-d.
Notes:
10.1.m
D3039
§493.1105
(a)(5)
Record Retention: Are quality
assessment records retained for at
least two years?
Note: Longer retention times might be
in place for other state, local, federal
regulations and must also be followed.
The CLIA two-year rule must be followed at
a minimum.
Records available for two years.
Note: Check state requirements as
applicable —go with most stringent.
Requirement met?
Yes
No
Notes:
10.1.n
D6094
§493.1445
(e)(5)
Technical Supervisor Responsibility
(Delegated by LD): Does the technical
supervisor ensure that QC and
QA programs are established and
maintained to assure the quality of
laboratory services provided and to
identify failures in quality as they occur?
QA policies/procedures and records.
Competency assessment of the
technical supervisor.
Requirement met?
Yes
No
See:
CLIA Personnel Review Worksheet Tool
Technical Supervisor Personnel Form
Notes:
SOM Appendix C (Rev. 166, 2017)
CFR 42, Part 493
APHL CLIA Audit Checklist | Quality Assessment & Corrective Action | Page 48
Table Key
Laboratory Director (LD)
Clinical Consultant (CC)
General Supervisor (GS)
Technical Supervisor (TS)
Laboratory Director: Date:
Return to
Table of Contents
10.2 Corrective Action
# SOM/CFR* Requirement Examples of Compliance Notes
10.2.a
D5779
§493.1282 (a)
Has the laboratory established
corrective action policies and
procedures?
Are the policies and procedures
available and followed in a manner that
ensures accurate and reliable patient
test results and reports?
Nonconforming event (NCE) and
corrective action procedure and records.
Requirement met?
Yes
No
Notes:
10.2.b
D5781
§493.1282
(b)(1)
Does the laboratory document
corrective actions taken when test
systems do not meet the laboratory’s
verified/established performance
specifications including:
Equipment or methodologies are
performed outside of established
operating parameters or performance
specifications?
Patient test values are outside of the
laboratory’s reportable range for the
test?
When the laboratory determines
that the reference intervals for a test
procedure are inappropriate for the
laboratory’s patient population?
NCE and corrective action reports show
evidence the laboratory is monitoring and
evaluating laboratory performance and
the quality of services.
NCE documentation should include all
specifics to be able to trace the event.
Example: Temperature-controlled spaces,
equipment and instruments must be
monitored and results documented
for acceptable temperature ranges.
If corrective action is needed when
acceptable temperature ranges are
exceeded, the serial number should be
included in the NCE.
Requirement met?
Yes
No
Notes:
10.2.c
D6178
§493.1495
(b)(4)
Testing Personnel Responsibility:
Is there documented evidence that
testing personnel follow the laboratory’s
established policies and procedures
whenever test systems are not within
established acceptable levels of
performance?
Documentation of corrective action or
NCE.
Requirement met?
Yes
No
See CLIA Personnel Review Worksheet Tool
Notes:
10.2.d
D6181
§493.1495
(b)(6)
Testing Personnel Responsibility:
Do testing personnel document
corrective actions taken when test
systems deviate from established
performance specifications?
Documentation of corrective action or
NCE.
Requirement met?
Yes
No
See CLIA Personnel Review Worksheet Tool
Notes:
10.2.e
D5783
§493.1282
(b)(2)
Does the laboratory document
corrective actions taken when results
of control or calibration materials
fail to meet established criteria for
acceptability including:
Evaluation of all patient results
obtained in the failed run and
previous runs since the most recent
acceptable run for acceptability?
The corrective action necessary to
ensure the reporting of accurate and
reliable patient test results?
NCE and corrective action procedure and
records include:
Retrospective review of patient results
from testing since the last time the
system was known to be performing
within specification
Issuing corrected patient results or
contacting physicians to notify of
potential error
Documentation of review with staff,
including signatures, action items,
follow-up checks for effectiveness and
evidence of trend monitoring.
Requirement met?
Yes
No
Notes:
Association of Public Health Laboratories
The Association of Public Health Laboratories (APHL) works to strengthen laboratory systems serving the public’s health
in the US and globally. APHL’s member laboratories protect the public’s health by monitoring and detecting infectious and
foodborne diseases, environmental contaminants, terrorist agents, genetic disorders in newborns and other diverse health
threats.
This publication was supported by Cooperative Agreements #NU60OE000103 and NU60OE000104, 100% funded by
the US Centers for Disease Control and Prevention (CDC). Its contents are solely the responsibility of the authors and do
not necessarily represent the ofcial views of CDC or the US Department of Health and Human Services.
8515 Georgia Avenue, Suite 700
Silver Spring, MD 20910
Phone: 240.485.2745
Fax: 240.485.2700
www.aphl.org
®
© Copyright 2022, Association of Public Health Laboratories. All Rights Reserved.