Application Bulletin 280/3 e
Automated water content determination with the 874 OSP
Page 2 of
Preparation
Sample
The sample is thoroughly mixed. Make sure the sample does
not change its water content during the mixing. The optimal
sample size depends on the water content of the sample.
Unfortunately there is no equation describing the relation
between water content and sample size. But there is a rule of
thumb: the higher the water content, the smaller the sample
size and vice versa. Sample weights that are too low have a
negative effect on the measuring accuracy (balance error).
The upper limit for the sample size is defined by the volume
of the sample vial. Theoretically the maximum sample weight
is also limited by the water capacity of the reagent. Usually
the capacity is approximately 1000 mg H
2
O per 100 mL
reagent (please contact reagent manufacturer for more
information). In other words, with one sample this capacity is
never used.
The absolute amount of water transferred to the titration
vessel is recommended to be in the range of 300 to 5000 µg.
If the absolute amount of water for a sample cannot be
reduced (e.g. smaller sample size …) also larger amounts of
water (> 5000 µg) can be determined using the oven
technique. Please be aware that in such cases the
determination times will increase.
The appropriate amounts of sample are weighed into the
sample vials and the vials are sealed with a septum cap.
Previous to usage, the vials and caps should be conditioned
for at least 24 hours at ambient air.
Table 2 on pages 5 and 6 shows recommended sample
weights for a choice of samples.
Instrument and software
The 874 Oven Sample Processor and the KF Titrator are
connected to a PC. The tiamo™ software is started. After the
instruments are recognized and therefore visible in the device
list, the work position for the conditioning vial, the tower
(sample positions) and the shift position are defined. The
tiamo™ software includes several preprogrammed methods
which can be loaded and used with an 874 system.
Depending on the KF Titrator, the methods need to be
adapted with the right instruments. For the analysis of sample
series, the following sequence of methods is recommended:
• systemprep
• blank value
• water content
The method “systemprep” is run once before a sample series.
This makes sure that the system of tubes is purged and ready
for the sample determinations. Afterwards the blank values of
empty sample vials are determined. Three replicas of the
blank value determination are recommended. Finally the
analysis of the water content of the sample takes place.
Normally a flow rate of the carrier gas (air, nitrogen or a
different inert gas) of 40 to 60 mL/min is sufficient. The flow
rate can be adjusted up to 150 mL/min to make sure that the
gas flow is high enough to transfer the released water as fast
as possible into the titration vessel. Generally the
determination time for liquid samples decreases if the gas
flow is increased. Avoid high gas flow rates when analyzing
solid samples which could swirl up.
The vials for conditioning, system preparation, determination
of the blank and the sample vials are placed on the rack of
the 874 Oven Sample Processor. For coulometric KF
titrations, the titration cell is filled with 150 mL of reagent and
then conditioned.
Temperature gradient
For samples whose temperature behavior is unknown, a so-
called temperature gradient is run (available temperature
range: 50 to 250 °C). The required method “temp gradient” is
a preprogrammed method included in the tiamo™ software.
This method uses a heating rate of 2 °C/min to heat up
samples from 50 to 250 °C. Figure 1 shows a theoretical
temperature gradient where the sample is heated from 50 to
250 °C in 100 min.
Fig. 1: Diagram showing a theoretical temperature gradient.
While a temperature gradient is being run it is possible to
record both, the amount of water released and the drift as a
function of time (see Fig. 2). The software tiamo™ offers the
possibility to correlate the amount of water released and the
drift with the oven temperature. This allows statements about
the kinetics of the water release as a function of the
temperature.